Chapter 3. Making Measurements
84
USM Go Operator’s Manual
3.8.1
Recording the DAC Curve
DAC Curve
points are typically recorded from a standard with equally-sized reflectors (holes) located at various material depths. The
primary echo from each of these points, for a total of up to 16 echoes, are recorded. When
DAC
mode is active, the instrument displays a
line that represents echo peaks for constant reflectors at varying material depth. This line is drawn as either a curved, linear in gain, line
based on a series of collected
DAC
points or a series of straight line segments that join these
DAC
points. Only one
DAC
curve can be
stored at a time. To program the
DAC
curve proceed as follows:
1.
In the
ACQUIRE
menu, activate the
DAC/TCG
submenu using the joystick (
).
2.
Couple the probe to the first reference point and use the
GATE A START
function to adjust Gate A so that it is broken by the primary
echo. If necessary, use the
Gain Toggle
to adjust the gain so that the echo crosses Gate A, and the highest peak in Gate A is at
approximately 80% of full-screen height. The highest peak must not be higher than 100% full-screen height.
Note:
The
AUTO 80
function allows for automatic application of gain to set the Gate A triggering echo to 80% of full screen height.
3.
While Gate A is lined up over the first reference echo, use the joystick (
) to select the function titled
RECORD
. Press the
center of the joystick (
) to activate the function. When the value of the
RECORD
function changes from 0 to 1, you have recorded
the first
DAC Curve
point, which is then treated as the reference echo. The amplitude value at which this point is recorded becomes the
“reference amplitude”
value. However, the
“reference amplitude”
is updated if a new point is recorded with an earlier
TOF
.
Содержание Ultrasonic USM Go
Страница 1: ...GE Sensing Inspection Technologies Ultrasonic P N 1259801 Rev 1 July 2009 USM Go Operator s Manual ...
Страница 2: ......
Страница 4: ... no content intended for this page proceed to next page ...
Страница 18: ...Preface xvi USM Go Operator s Manual no content intended for this page proceed to next page ...
Страница 80: ...Chapter 2 Instrument Setup 62 USM Go Operator s Manual no content intended for this page proceed to next page ...
Страница 130: ...Chapter 3 Making Measurements 112 USM Go Operator s Manual no content intended for this page proceed to next page ...
Страница 152: ...Chapter 5 Data Recorder Files 134 USM Go Operator s Manual no content intended for this page proceed to next page ...
Страница 162: ...144 USM Go Operator s Manual no content intended for this page proceed to next page ...
Страница 178: ...Appendix C EN 12668 Specifications 160 USM Go Operator s Manual ...
Страница 189: ......