Chapter 2. Instrument Setup
USM Go Operator’s Manual
49
2.4.8 Selecting a Rectification Mode (cont.)
1.
In the
ACQUIRE
menu, activate the
RECEIVER
submenu using the joystick (
).
2.
Use the joystick (
) to select the function titled
RECTIFY
. The following options are available:
•
NEG HALFWAVE
-
Shows the negative component of the RF signal but displays it in a positive orientation.
•
POS HALFWAVE
-
Shows the positive component of the RF signal.
•
FULLWAVE
-
Shows the positive and negative halves of the RF wave, but both are oriented in the positive direction.
•
RF
-
Shows the echo with no rectification.
3.
Press the function toggle or move the joystick (
) to select the desired rectification method.
4.
After completing your selection, move the joystick up or down (
) to navigate away from this function.
Содержание Ultrasonic USM Go
Страница 1: ...GE Sensing Inspection Technologies Ultrasonic P N 1259801 Rev 1 July 2009 USM Go Operator s Manual ...
Страница 2: ......
Страница 4: ... no content intended for this page proceed to next page ...
Страница 18: ...Preface xvi USM Go Operator s Manual no content intended for this page proceed to next page ...
Страница 80: ...Chapter 2 Instrument Setup 62 USM Go Operator s Manual no content intended for this page proceed to next page ...
Страница 130: ...Chapter 3 Making Measurements 112 USM Go Operator s Manual no content intended for this page proceed to next page ...
Страница 152: ...Chapter 5 Data Recorder Files 134 USM Go Operator s Manual no content intended for this page proceed to next page ...
Страница 162: ...144 USM Go Operator s Manual no content intended for this page proceed to next page ...
Страница 178: ...Appendix C EN 12668 Specifications 160 USM Go Operator s Manual ...
Страница 189: ......