USM Go Operator’s Manual
137
A.3 Pulser
Note:
All pulser measurements were taken according to EN12668 specifications.
Pulser Mode:
Standard:
simulated spike
Optional:
uni-polar square wave via software control
Pulser Voltage (SQ Mode):
20 V to 300 V, with a 10 V step and tolerance of 10%
Pulser Falling/Rising Time:
10 ns maximum
Pulser Width (SQ Mode):
30 ns to 500 ns, with a 20 ns step and a tolerance of 10%
Pulser Amplitude (Spike Mode):
Low:
120 V
High:
300 V
Damping:
50 ohms or 1000 ohms