Chapter 2. Instrument Setup
50
USM Go Operator’s Manual
2.4.9 Setting the A-Scan REJECT Level
A portion of the A-Scan can be omitted from the display screen, by defining the percentage of the full-screen height you wish to omit. To set
a reject percentage:
1.
In the
ACQUIRE
menu, activate the
RECEIVER
submenu using the joystick (
).
2.
Use the joystick (
) to select the function titled
REJECT
.
3.
To change the amount of the A-Scan you wish to omit from the display, as a percentage of the full-screen height, either move the
joystick (
) or press the
function toggle
. You may omit a portion of the A-Scan up to 80% of the full-screen height.
Note:
Whenever
REJECT
is set to a value greater then 0%, the reject icon (see
Figure 6 on page 17
) is displayed in the status bar.
Содержание Ultrasonic USM Go
Страница 1: ...GE Sensing Inspection Technologies Ultrasonic P N 1259801 Rev 1 July 2009 USM Go Operator s Manual ...
Страница 2: ......
Страница 4: ... no content intended for this page proceed to next page ...
Страница 18: ...Preface xvi USM Go Operator s Manual no content intended for this page proceed to next page ...
Страница 80: ...Chapter 2 Instrument Setup 62 USM Go Operator s Manual no content intended for this page proceed to next page ...
Страница 130: ...Chapter 3 Making Measurements 112 USM Go Operator s Manual no content intended for this page proceed to next page ...
Страница 152: ...Chapter 5 Data Recorder Files 134 USM Go Operator s Manual no content intended for this page proceed to next page ...
Страница 162: ...144 USM Go Operator s Manual no content intended for this page proceed to next page ...
Страница 178: ...Appendix C EN 12668 Specifications 160 USM Go Operator s Manual ...
Страница 189: ......