Chapter 3. Making Measurements
90
USM Go Operator’s Manual
3.10.2 Setting Transfer Correction (DAC/TCG-MAT ATTN-TRANSFER CORR)
To adjust the dB compensation for differences in ultrasonic coupling between the standard and the test piece:
1.
Access the
MAT ATTN
submenu in the
DAC/TCG
menu.
2.
Adjust the
TRANSFER CORR
function as required to compensate for coupling differences.
Note:
When
TRANSFER CORR
is set to any value other than 0, the displayed gain value appears in a contrasting color, indicating that
the amount of applied gain differs from the listed
“instrument”
gain.
Содержание Ultrasonic USM Go
Страница 1: ...GE Sensing Inspection Technologies Ultrasonic P N 1259801 Rev 1 July 2009 USM Go Operator s Manual ...
Страница 2: ......
Страница 4: ... no content intended for this page proceed to next page ...
Страница 18: ...Preface xvi USM Go Operator s Manual no content intended for this page proceed to next page ...
Страница 80: ...Chapter 2 Instrument Setup 62 USM Go Operator s Manual no content intended for this page proceed to next page ...
Страница 130: ...Chapter 3 Making Measurements 112 USM Go Operator s Manual no content intended for this page proceed to next page ...
Страница 152: ...Chapter 5 Data Recorder Files 134 USM Go Operator s Manual no content intended for this page proceed to next page ...
Страница 162: ...144 USM Go Operator s Manual no content intended for this page proceed to next page ...
Страница 178: ...Appendix C EN 12668 Specifications 160 USM Go Operator s Manual ...
Страница 189: ......