![GE Ultrasonic USM Go Скачать руководство пользователя страница 174](http://html.mh-extra.com/html/ge/ultrasonic-usm-go/ultrasonic-usm-go_operators-manual_128490174.webp)
Appendix C. EN 12668 Specifications
156
USM Go Operator’s Manual
Paragraph Parameter
LSL Typ.
USL
Units Test
Conditions
9.5.3 Equivalent
Input
Noise
-
50
80
nV/
sqrt Hz
Low Pass Selected
-
37
80
nV/
sqrt Hz
4 to 5 MHz Selected
-
33
80
nV/
sqrt Hz
10 MHz Selected
-
40
80
nV/
sqrt Hz
13 MHz Selected
-
45
80
nV/
sqrt Hz
BroadBand Selected
9.5.4
Accuracy of Calibrated Attenuator
-1
0.7
1
dB
Cumulative Error over 20dB Range
-2
0.7
2
dB
Cumulative Error over 60dB Range
Table 2: EN 12668-1:2000 Specifications (cont.)
Содержание Ultrasonic USM Go
Страница 1: ...GE Sensing Inspection Technologies Ultrasonic P N 1259801 Rev 1 July 2009 USM Go Operator s Manual ...
Страница 2: ......
Страница 4: ... no content intended for this page proceed to next page ...
Страница 18: ...Preface xvi USM Go Operator s Manual no content intended for this page proceed to next page ...
Страница 80: ...Chapter 2 Instrument Setup 62 USM Go Operator s Manual no content intended for this page proceed to next page ...
Страница 130: ...Chapter 3 Making Measurements 112 USM Go Operator s Manual no content intended for this page proceed to next page ...
Страница 152: ...Chapter 5 Data Recorder Files 134 USM Go Operator s Manual no content intended for this page proceed to next page ...
Страница 162: ...144 USM Go Operator s Manual no content intended for this page proceed to next page ...
Страница 178: ...Appendix C EN 12668 Specifications 160 USM Go Operator s Manual ...
Страница 189: ......