Chapter 3. Making Measurements
88
USM Go Operator’s Manual
3.9.2
Working with TCG
In TCG mode the instrument uses the recorded reference points to calculate an amount of gain correction required to display each echo from
same-size reflectors at the same amplitude. The recorded reference point data is stored until replaced or edited. To use the stored reference
points and operate in
TCG
mode:
1.
Enter the
SETUP
menu, and then select the
TCG/DAC MODE
function.
2.
Activate the
TCG/DAC MODE
function, and select the
TCG
option.
3.
Select the
TCG DISPLAY
function, and turn this feature
ON
or
OFF
.
Note:
The TCG curve begins at the first reference point recorded. It then proceeds horizontally from the amplitude of this first reference
point to the depth (i.e. time position) of the last reference point recorded.
Note:
When rectification is set to
RF
, reference lines are not displayed.
Содержание Ultrasonic USM Go
Страница 1: ...GE Sensing Inspection Technologies Ultrasonic P N 1259801 Rev 1 July 2009 USM Go Operator s Manual ...
Страница 2: ......
Страница 4: ... no content intended for this page proceed to next page ...
Страница 18: ...Preface xvi USM Go Operator s Manual no content intended for this page proceed to next page ...
Страница 80: ...Chapter 2 Instrument Setup 62 USM Go Operator s Manual no content intended for this page proceed to next page ...
Страница 130: ...Chapter 3 Making Measurements 112 USM Go Operator s Manual no content intended for this page proceed to next page ...
Страница 152: ...Chapter 5 Data Recorder Files 134 USM Go Operator s Manual no content intended for this page proceed to next page ...
Страница 162: ...144 USM Go Operator s Manual no content intended for this page proceed to next page ...
Страница 178: ...Appendix C EN 12668 Specifications 160 USM Go Operator s Manual ...
Страница 189: ......