Chapter 2. Instrument Setup
USM Go Operator’s Manual
39
2.4.2b Specifying the Probe Frequency (
RECEIVER-FREQUENCY)
1.
In the
ACQUIRE
menu, activate the
RECEIVER
submenu using the joystick (
).
2.
Use the joystick (
) to select the function titled
FREQUENCY
.
3.
To change the specified frequency, move the joystick (
) or press the function toggle. The following options are available:
•
Low Pass, 4, 5, 10, 13 MHz
-
Choose the frequency that matches the frequency of your probe.
•
BROADBAND
-
Select this option to use the built-in broadband filter.
4.
After completing your selection, move the joystick up or down (
) to navigate away from this function.
Содержание Ultrasonic USM Go
Страница 1: ...GE Sensing Inspection Technologies Ultrasonic P N 1259801 Rev 1 July 2009 USM Go Operator s Manual ...
Страница 2: ......
Страница 4: ... no content intended for this page proceed to next page ...
Страница 18: ...Preface xvi USM Go Operator s Manual no content intended for this page proceed to next page ...
Страница 80: ...Chapter 2 Instrument Setup 62 USM Go Operator s Manual no content intended for this page proceed to next page ...
Страница 130: ...Chapter 3 Making Measurements 112 USM Go Operator s Manual no content intended for this page proceed to next page ...
Страница 152: ...Chapter 5 Data Recorder Files 134 USM Go Operator s Manual no content intended for this page proceed to next page ...
Страница 162: ...144 USM Go Operator s Manual no content intended for this page proceed to next page ...
Страница 178: ...Appendix C EN 12668 Specifications 160 USM Go Operator s Manual ...
Страница 189: ......