Chapter 5. Troubleshooting and Repair
122
PanaFlow™ MV80 & MV82 User’s Manual
5.4.5 Symptom: Meter Displays Pressure Fault
1.
For remote mounted electronics, carefully check all the wiring connections in the remote mount junction box.
There are 18 connections that must be correct. Verify each color (black and red), shield, and wire number.
2.
Using ESD precautions and hazardous area precautions, remove the electronics enclosure window cover.
Disconnect the pressure sensor from the electronics stack (see
) or the remote feed
through board (see
). Measure the resistance across the outside pins of the pressure
sensor connector, then across the inside pins (see
below). Both readings should be approximately
4000 ohms.
Figure 82: Pressure Sensor Connector
3.
Go to the first column of the hidden diagnostics and record the
Pe(V)
and
Pv(V)
values and consult GE with your
findings.
OUTSIDE
INSIDE
OUTSIDE
INSIDE
Содержание PanaFlow MV80
Страница 2: ......
Страница 12: ...Contents x PanaFlow MV80 MV82 User s Manual no content intended for this page ...
Страница 20: ...Chapter 1 Introduction 8 PanaFlow MV80 MV82 User s Manual no content intended for this page ...
Страница 62: ...Chapter 2 Installation 50 PanaFlow MV80 MV82 User s Manual no content intended for this page ...
Страница 120: ...Chapter 4 Serial Communications 108 PanaFlow MV80 MV82 User s Manual no content intended for this page ...
Страница 148: ...Appendix B Approvals 136 PanaFlow MV80 MV82 User s Manual no content intended for this page ...
Страница 158: ...Appendix D Glossary 146 PanaFlow MV80 MV82 User s Manual no content intended for this page ...
Страница 160: ...Warranty 148 PanaFlow MV80 MV82 User s Manual no content intended for this page ...
Страница 162: ... no content intended for this page ...
Страница 163: ......