PanaFlow™ MV80 & MV82 User’s Manual
119
Chapter 5. Troubleshooting and Repair
5.4.3 Symptom: No Output
1.
For remote mounted electronics, carefully check all the wiring connections in the remote mount junction box.
There are 18 connections that must be correct, verify each color (black and red), shield, and wire number.
2.
Turn on the pressure and temperature display in the
Display Menu
and verify that the pressure and temperature
are correct.
3.
Using ESD precautions and hazardous area precautions, remove the electronics enclosure window cover.
Disconnect the vortex sensor from the electronics stack (see
below) or the remote feed through
below). Measure the resistance from each outside pin to the meter ground (see
) - each should be open. Measure the resistance from the center pin to the meter ground - this
should be grounded to the meter.
Figure 78: Electronics Stack Sensor Connections
Figure 79: Remote Feed Through Board Sensor Connections
VORTEX
TEMPERATURE
PRESSURE
VORTEX
TEMPERATURE
PRESSURE
VORTEX
PRESSURE
TEMPERATURE
Содержание PanaFlow MV80
Страница 2: ......
Страница 12: ...Contents x PanaFlow MV80 MV82 User s Manual no content intended for this page ...
Страница 20: ...Chapter 1 Introduction 8 PanaFlow MV80 MV82 User s Manual no content intended for this page ...
Страница 62: ...Chapter 2 Installation 50 PanaFlow MV80 MV82 User s Manual no content intended for this page ...
Страница 120: ...Chapter 4 Serial Communications 108 PanaFlow MV80 MV82 User s Manual no content intended for this page ...
Страница 148: ...Appendix B Approvals 136 PanaFlow MV80 MV82 User s Manual no content intended for this page ...
Страница 158: ...Appendix D Glossary 146 PanaFlow MV80 MV82 User s Manual no content intended for this page ...
Страница 160: ...Warranty 148 PanaFlow MV80 MV82 User s Manual no content intended for this page ...
Страница 162: ... no content intended for this page ...
Страница 163: ......