Chapter 5. Troubleshooting and Repair
120
PanaFlow™ MV80 & MV82 User’s Manual
5.4.3 Symptom: No Output (cont.)
Figure 80: Vortex Sensor Connector Pins
4.
With the sensor still disconnected, go to the first column of the hidden diagnostics and display the vortex
shedding frequency,
f
. Hold a finger on the three exposed pins on the analog board. The meter should read
electrical noise, 60 Hz for example. If all readings are correct, re-install the vortex sensor wires.
5.
Verify all meter configuration and troubleshooting steps previously described. There are many possible causes
of this problem. Consult GE if necessary.