Chapter 5. Troubleshooting and Repair
118
PanaFlow™ MV80 & MV82 User’s Manual
5.4 Determine the Fault
5.4.1 Symptom: Output at No Flow
•
The low flow cutoff is set too low. At no flow, go to the first column of the hidden diagnostics menu and record
the
Lvl
value. The low flow cutoff must be set above this value.
•
Example: at no flow,
Lvl
= 25. Set the low flow cutoff in the Calibration Menu to approximately 28 and the
meter will no longer read a flow rate at no flow.
5.4.2 Symptom: Erratic Output
•
The flow rate may be too low, just at the cutoff of the meter range, and the flow cycles above and below the
cutoff making an erratic output. Consult the factory if necessary to confirm the meter range based on current
operating conditions. It may be possible to lower the low flow cutoff to increase the meter range. See the
example above for output at no flow, only this time the low flow cutoff is set too high. You can lower this value
to increase the meter range as long as you do not create the output at no flow condition previously described.
•
Mechanical installation may be incorrect. Verify the straight run is adequate, as described in
Chapter 2
. For
in-line meters, make sure the meter is not installed backwards and there are no gaskets protruding into the flow
stream. For insertion meters, verify the insertion depth and flow direction.
•
The meter may be reacting to actual changes in the flow stream. The output can be smoothed using a time
constant. The displayed values can be smoothed using the time constant in the
Display Menu
. The analog
outputs can be smoothed using the time constant in the
Output Menu
. A time constant of 1 will result in the
change in value reaching 63% of its final value in one second. A time constant of 4 is 22%, 10 is 9.5% and 50
is 1.9% of the final value in one second. The time constant equation is shown below (TC = Time Constant).
•
The vortex coefficient
Ck
may be incorrectly set. The
Ck
is a value in the equation used to determine if a
frequency represents a valid vortex signal given the fluid density and signal amplitude. In practice, the
Ck
value
controls the adaptive filter,
fi
, setting. During flow, view the
f
and
fi
values in the first column of the hidden
diagnostics. The
fi
value should be approximately 10-20% higher than the
f
value. If you raise the
Ck
setting in
the
Calibration Menu
, then the
fi
value will increase. The
fi
is a low pass filter, so by increasing it or lowering
it, you can alter the range of frequencies that the meter will accept. If the vortex signal is strong, the
fi
value
will increase to a large number - this is correct.
% change in final value in one second
100 1 e
-1 TC
⁄
–
(
)
=
Содержание PanaFlow MV80
Страница 2: ......
Страница 12: ...Contents x PanaFlow MV80 MV82 User s Manual no content intended for this page ...
Страница 20: ...Chapter 1 Introduction 8 PanaFlow MV80 MV82 User s Manual no content intended for this page ...
Страница 62: ...Chapter 2 Installation 50 PanaFlow MV80 MV82 User s Manual no content intended for this page ...
Страница 120: ...Chapter 4 Serial Communications 108 PanaFlow MV80 MV82 User s Manual no content intended for this page ...
Страница 148: ...Appendix B Approvals 136 PanaFlow MV80 MV82 User s Manual no content intended for this page ...
Страница 158: ...Appendix D Glossary 146 PanaFlow MV80 MV82 User s Manual no content intended for this page ...
Страница 160: ...Warranty 148 PanaFlow MV80 MV82 User s Manual no content intended for this page ...
Страница 162: ... no content intended for this page ...
Страница 163: ......