PanaFlow™ MV80 & MV82 User’s Manual
137
Appendix C. Flowmeter Calculations
Appendix C. Flowmeter Calculations
C.1 In-Line Flowmeter Calculations
C.1.1 Volume Flow Rate
C.1.2 Mass Flow Rate
C.1.3 Flowing Velocity
Where:
A
= Cross sectional area of the pipe (ft
2
)
f
= Vortex shedding frequency (pulses/sec)
K
= Meter factor corrected for thermal expansion (pulses/ft
3
)
Q
M
= Mass flow rate (lbm/sec)
Q
V
= Volume flow rate (ft
3
/sec)
V
f
= Flowing velocity (ft/sec)
ρ
= Density (lbm/ft
3
)
Qv
f
K
---
=
Q
M
Q
V
ρ
=
V
f
Q
V
A
-----
=
Содержание PanaFlow MV80
Страница 2: ......
Страница 12: ...Contents x PanaFlow MV80 MV82 User s Manual no content intended for this page ...
Страница 20: ...Chapter 1 Introduction 8 PanaFlow MV80 MV82 User s Manual no content intended for this page ...
Страница 62: ...Chapter 2 Installation 50 PanaFlow MV80 MV82 User s Manual no content intended for this page ...
Страница 120: ...Chapter 4 Serial Communications 108 PanaFlow MV80 MV82 User s Manual no content intended for this page ...
Страница 148: ...Appendix B Approvals 136 PanaFlow MV80 MV82 User s Manual no content intended for this page ...
Страница 158: ...Appendix D Glossary 146 PanaFlow MV80 MV82 User s Manual no content intended for this page ...
Страница 160: ...Warranty 148 PanaFlow MV80 MV82 User s Manual no content intended for this page ...
Страница 162: ... no content intended for this page ...
Страница 163: ......