PanaFlow™ MV80 & MV82 User’s Manual
21
Chapter 2. Installation
2.3.2 Installing Flowmeters with a Compression Connection (cont.)
Example:
To install a Series MV82 meter with a standard probe (S = 29.47 inches) into a 14 inch schedule 40 pipe, the following
measurements (in inches) are taken:
•
F = 3 inches
•
R = 5 inches
•
t = 0.438 inches
Then, using the formula given in
, the insertion length for this example is 21.03 inches. Insert the
stem through the fitting until an insertion length of 21.03 inches is measured with a ruler.
Figure 12: Flowmeter with a Compression Type Fitting
Compression Nut
Stem Housing
2-inch NPT
Connection
Compression Nut
Sensor Head
Flange
Connection
Sensor Alignment
Pointer
Encloser Adapter
Stem
Stem Housing