Sample collection using the Vital Capacity Bag kit consists of three steps:
•
Connecting a pre-cleaned bag to the filler.
•
Instructing the subject on how to perform the maneuver and collecting the
samples
•
Disconnecting the bag from the filler and sealing the bag.
Connecting the bag to the filler
To prepare the filler for sample collection:
•
Stand the filler on the inspiratory filter.
•
Remove the cap from the bag and grasp the cap, the top of the bag and
internal tube with one hand while holding the bag filler with the other hand.
•
Twist the bag slightly counterclockwise, then insert the Luer fitting on the bag
into the Luer fitting on the filler and rotate fully clockwise until the o-ring
seals against the base of the bag filler.
•
Rotate the bag so that there are no folds or restrictions at the end of the tube.
•
Open a new viral and bacterial filter and connect the filter to the tapered end
on the top of the filler.
•
Connect a sterile mouthpiece to the filter.
Instructing the Subject and Collecting the Samples
To perform the maneuver:
•
Have the subject comfortably seated and holding the bag filler level.
•
Have the subject place the mouthpiece in their mouth.
•
Instruct the subject to inhale orally through the inspiratory filter to total lung
capacity.
•
Have the subject immediately exhale while watching the pressure gauge and
adjusting the force of exhalation to maintain a pressure of 13 cm H
2
O to give
an expiratory flow rate of 0.35 L/s.* During the exhalation, the pressure should
* at 760 torr, for other pressures, calculate the required pressure from the equation:
Содержание MOA 280i
Страница 6: ...GE Analytical Instruments 2006 DLM 14291 Rev A vi ...
Страница 14: ...GE Analytical Instruments 2006 DLM 14291 Rev A xiv ...
Страница 24: ......
Страница 42: ...GE Analytical Instruments 2006 DLM 14291 Rev A 4 8 Figure 4 3 NOA Back Panel ...
Страница 52: ......
Страница 72: ......
Страница 84: ......
Страница 118: ...GE Analytical Instruments 2006 11 6 DLM 14291 Rev A ...
Страница 124: ......
Страница 142: ......
Страница 160: ...GE Analytical Instruments 2006 13 18 DLM 14291 Rev A ...
Страница 168: ......
Страница 184: ......
Страница 200: ...GE Analytical Instruments 2006 16 16 DLM 14291 Rev A ...