GE Analytical Instruments ©2006
DLM 14291 Rev. A
3-3
Figure 3-5:Measurement Menu (Nitric Oxide Mode)
V A C U U M
P U M P
P A S S
C O O L E R
T E M P
P A S S
C E L L P R E S S
P A S S
S U P P L Y
P R E S S
T E S T I N G
P M T S I G N A L
E N T = S T A T U S
C L R = S T O P
Figure 3-4: Start-up Screen.
•
Check the PMT cooler temperature to see if it is –12 ± 2 °C.
•
Wait until the reaction cell pressure is < 100 torr.
•
Check the ozone supply pressure is >4 psig (6 psig recommended).
•
Turn on the PMT and record an ozone-off baseline signal.
•
Turn on the ozone generator, and wait for an increase in the PMT signal
due to the background chemiluminescence from ozone.
•
Display the PMT signal in the Measurement Display.
If any of the above conditions are not met, the start-up screen will FAILED for
that test. Pressing ENTER from the Start-up Screen will display the Status
Screen to aid in troubleshooting failed tests. Pressing CLEAR will return to the
Analysis Menu.
Measurement Menu
Once the start-up testing is completed, the Measurement Menu is displayed.
Содержание MOA 280i
Страница 6: ...GE Analytical Instruments 2006 DLM 14291 Rev A vi ...
Страница 14: ...GE Analytical Instruments 2006 DLM 14291 Rev A xiv ...
Страница 24: ......
Страница 42: ...GE Analytical Instruments 2006 DLM 14291 Rev A 4 8 Figure 4 3 NOA Back Panel ...
Страница 52: ......
Страница 72: ......
Страница 84: ......
Страница 118: ...GE Analytical Instruments 2006 11 6 DLM 14291 Rev A ...
Страница 124: ......
Страница 142: ......
Страница 160: ...GE Analytical Instruments 2006 13 18 DLM 14291 Rev A ...
Страница 168: ......
Страница 184: ......
Страница 200: ...GE Analytical Instruments 2006 16 16 DLM 14291 Rev A ...