GE Analytical Instruments ©2006
15-9
DLM 14291 Rev. A
HIGH VOLTAGE
to power supply
SIGNAL
to amplifier
Figure 15-4: Connections to PMT
•
If the vacuum hose was removed from the exhaust of the NOA, replace the hose
and tighten with a wrench.
•
Turn the room lights off, remove the PMT from the drawer and remove the
black cap being careful not to touch the face of the PMT. If the face of the PMT
is touched, use a clean, dry cloth to remove any fingerprints.
•
Insert the PMT into the cooler and reinstall the socket by turning the socket
clockwise. Tighten the socket until it can no longer be turned by hand.
•
Reconnect the high voltage and signal cables to the PMT. The connectors are
different sizes and the cables can only be connected one way.
•
Position the cover for the NOA so that the slots on the side of the cover are
pointing towards to back of the NOA. Place the cover over the NOA, being
careful to clear all of the wires on both sides of the analyzer. Position the back
of the cover on top of the back
panel of the NOA. Position the front
of the cover so that it rests on the
front panel of the NOA. Move the
front and rear panels a little to help align the cover.
•
With the cover resting on top of both the front and rear panels, position the
front panel to align the front, right-hand screw holes in the front of the
analyzer and install a screw a few turns. Repeat the process to align the top
right-hand screw at the back of the analyzer and partially install a screw.
•
Repeat this procedure to align the holes on the top, left-hand side of the
analyzer (front and back) and partially install the screws.
•
Align the bottom screw holes and partially install the screws.
•
Repeat the process until all screws are partially installed. If the screw holes do
not all line-up, the cover has been installed backwards. Remove all of the
screws, turn the cover around and repeat the process.
•
After all of the screws have been installed, use a phillips-head screwdriver to
tighten the cover.
•
Reconnect the power cord and turn the main power switch on.
Содержание MOA 280i
Страница 6: ...GE Analytical Instruments 2006 DLM 14291 Rev A vi ...
Страница 14: ...GE Analytical Instruments 2006 DLM 14291 Rev A xiv ...
Страница 24: ......
Страница 42: ...GE Analytical Instruments 2006 DLM 14291 Rev A 4 8 Figure 4 3 NOA Back Panel ...
Страница 52: ......
Страница 72: ......
Страница 84: ......
Страница 118: ...GE Analytical Instruments 2006 11 6 DLM 14291 Rev A ...
Страница 124: ......
Страница 142: ......
Страница 160: ...GE Analytical Instruments 2006 13 18 DLM 14291 Rev A ...
Страница 168: ......
Страница 184: ......
Страница 200: ...GE Analytical Instruments 2006 16 16 DLM 14291 Rev A ...