GE Analytical Instruments ©2006
DLM 14291 Rev. A
3-6
C O N T R O L
M E T H O D
S E T U P
S T A T U S
Figure 3-8: Control Menu.
A N A L Y S I S
S T O P
S T A N D - B Y
Figure 3-7: Analysis Menu (NOA in Start Mode).
When the NOA is in the start mode, selecting Stop will:
•
Display a confirmation screen “Are you sure? continue / stop”
•
If the Stop option is selected, turn off the PMT, and ozone generator.
•
Run the vacuum pump for 2 minutes to clear residual ozone from the
system.
•
Turn off the vacuum pump.
Main Menu Options
The other options from the Main Menu are used to configure the NOA for the
different applications, perform the calibration for gas-phase measurements,
view warnings and errors, and view and install consumables.
Control
The Control Menu is used to setup the NOA for liquid or gas-phase
measurements, set the output intervals and the sensitivity of the NOA. The
setup can be stored as a method to permit easy switching between
applications. The Control Menu also permits viewing of the cooler temperature
Содержание MOA 280i
Страница 6: ...GE Analytical Instruments 2006 DLM 14291 Rev A vi ...
Страница 14: ...GE Analytical Instruments 2006 DLM 14291 Rev A xiv ...
Страница 24: ......
Страница 42: ...GE Analytical Instruments 2006 DLM 14291 Rev A 4 8 Figure 4 3 NOA Back Panel ...
Страница 52: ......
Страница 72: ......
Страница 84: ......
Страница 118: ...GE Analytical Instruments 2006 11 6 DLM 14291 Rev A ...
Страница 124: ......
Страница 142: ......
Страница 160: ...GE Analytical Instruments 2006 13 18 DLM 14291 Rev A ...
Страница 168: ......
Страница 184: ......
Страница 200: ...GE Analytical Instruments 2006 16 16 DLM 14291 Rev A ...