GE Analytical Instruments ©2006
7-7
DML 14291 Rev. A
Figure 7-6: Flow/Pressure Calibrations for Restrictors
12) in which the difference between the [NO] at the beginning and the [NO] at the
end of the interval varies by < 10% (based on the lowest [NO]). At no time during
the interval should the [NO] exceed the maximum NO at the beginning or end. For
exhaled NO levels < 5 ppb, the beginning and end NO values should not differ by
more than 1 ppb.
The REB program uses these criteria to select the value for the NO plateau. Three
exhalation profiles can be observed; constant NO versus time, increasing NO versus
time and decreasing NO versus time.
For increasing NO plateaus, the end [NO] minus the beginning [NO] divided by the
beginning [NO] must be < 10%, the [NO] during the interval must be less than the
end [NO] and the end [NO] is selected as the exhaled NO value.
For decreasing NO plateaus, the beginning [NO] minus the end [NO] divided by the
end [NO] must be < 10%, the [NO] during the interval must be less than the
Содержание MOA 280i
Страница 6: ...GE Analytical Instruments 2006 DLM 14291 Rev A vi ...
Страница 14: ...GE Analytical Instruments 2006 DLM 14291 Rev A xiv ...
Страница 24: ......
Страница 42: ...GE Analytical Instruments 2006 DLM 14291 Rev A 4 8 Figure 4 3 NOA Back Panel ...
Страница 52: ......
Страница 72: ......
Страница 84: ......
Страница 118: ...GE Analytical Instruments 2006 11 6 DLM 14291 Rev A ...
Страница 124: ......
Страница 142: ......
Страница 160: ...GE Analytical Instruments 2006 13 18 DLM 14291 Rev A ...
Страница 168: ......
Страница 184: ......
Страница 200: ...GE Analytical Instruments 2006 16 16 DLM 14291 Rev A ...