![GE MOA 280i Скачать руководство пользователя страница 139](http://html.mh-extra.com/html/ge/moa-280i/moa-280i_operation-and-maintenance-manual_128401139.webp)
GE Analytical Instruments ©2006
12-15
DLM 14291 Rev. A
Figure 12-3: Cleaning the Purge Vessel Frit.
•
If the condenser cannot be totally cleaned by rinsing, loosen the clamp so that
the purge vessel can be tilted. Fill the purge vessel with water then replace the
screw cap a few turns so that it is not tightly sealed, and will allow air to
escape. Tilt the purge vessel to fill of the condenser with water then return to
the upright position. Repeat until all of the contamination has been removed.
•
To clean the frit, turn off the gas inlet stopcock and unscrew and remove the
needle valve on the purge vessel. Use a wash bottle or pipette to fill the glass
tube between the needle valve connection and the frit with water.
•
Replace the needle valve, and open the gas inlet stopcock to allow the gas to
force the water through the frit. Once the water has been passed through the
frit, turn off the gas inlet stopcock, remove the needle valve, and add more
water to the tube. Replace the needle valve, and turn on the gas to force the
water through the frit. Repeat this procedure 3 or 4 times to completely clean
the frit and tubing.
•
After cleaning the frit, replace the reducing union on the outlet of the purge
vessel. Tighten the 1/4" nut fingertight. Do not use a wrench to tighten the nut
because the glass will break if the nut is overtightened (see Procedure for
Tightening Swagelok Fittings page 11-4).
Содержание MOA 280i
Страница 6: ...GE Analytical Instruments 2006 DLM 14291 Rev A vi ...
Страница 14: ...GE Analytical Instruments 2006 DLM 14291 Rev A xiv ...
Страница 24: ......
Страница 42: ...GE Analytical Instruments 2006 DLM 14291 Rev A 4 8 Figure 4 3 NOA Back Panel ...
Страница 52: ......
Страница 72: ......
Страница 84: ......
Страница 118: ...GE Analytical Instruments 2006 11 6 DLM 14291 Rev A ...
Страница 124: ......
Страница 142: ......
Страница 160: ...GE Analytical Instruments 2006 13 18 DLM 14291 Rev A ...
Страница 168: ......
Страница 184: ......
Страница 200: ...GE Analytical Instruments 2006 16 16 DLM 14291 Rev A ...