3. Setting Up the DMS 2
Page 84
DMS 2 Operating Manual
Step 2: Make sure the data file that you wish to edit is
active, and the specific point (or a point within the elevation
or location you wish to rename) is selected.
Step 3: Activate the EDIT submenu by selecting it with
or
so that it’s highlighted. Several selections will
appear, including ITEM.
Step 4: Press
below the selection titled ITEM. You’ll
note that several file characteristics are listed including some
of the following:
Point, Row, Column, Elevation, or Location.
This listing varies depending on the selected file type. Not all
ITEM’s listed can be modified.
Step 5: Select the item you wish to modify. For instance, if
you wish to change the name given to an Elevation (like the
elevation name shown in Figure 3-11) from ELEV1 to
LEVEL1, you must first press
below the ITEM selection
until ELEVATION is highlighted.
Step 6: Then use the AT selection to move to the desired
location in the data file. In this case, press
below AT
until ELEV1 appears next to the word ELEVATION. This also
causes the specified data point stored in ELEV1 to become
active in the File NAV Window. (
Note: You must select
insert or delete mode for AT to be available.)
Step 7: Press
to open the Virtual Keyboard. Use the
keyboard (as explained in Section 3.4.2) to specify a new
Elevation name, then press
to return to the DR Mode.
Step 8: The label selected (in this case the Elevation name)
will be set to the value specified.
3.4.5
Storing A-scans with Thickness Measurements
Attaching A-scans to Data Points — The SEND Button’s
Function
Pressing
normally results in a thickness measure-
ment being stored in the active file position. It is also pos-
sible to store an A-scan in the active file position, together
with the thickness measurement, by adjusting the value of
Содержание MFE DMS 2
Страница 2: ...DMS 2 DMS 2E Operating Manual GE Inspection Technologies Ultrasonics Ident Nr 021 002 151 Rev H ...
Страница 3: ...THIS PAGE INTENTIONALLY BLANK ...
Страница 22: ...1 Getting Started DMS 2 Operating Manual Page 11 FIGURE 1 7 The VIRTUAL KEYBOARD is shown here ...
Страница 94: ...3 Setting Up the DMS 2 DMS 2 Operating Manual Page 83 FIGURE 3 11 An Auto Labeler screen is shown here ...