5. Working with Stored Data
Page 140
DMS 2 Operating Manual
5.1.1
Selecting an Existing Data File
Use this procedure to select an existing data file. Selecting a
file automatically opens it. Note that a file can only be selected
from the Data Recorder Menu. (To create a new data file, refer
to Section 3.4.3.)
Step 1: Open the Data Recorder Menu accessed from the
Test or TG Mode by pushing
two times. Once the DR
Menu (Figure 5-2) is opened, you’ll note five submenus listed
on the Submenu Bar and three to five selections listed across
the bottom of the display.
Step 2: Activate any submenu located in the DR Menu by
selecting it with
or
so that it’s highlighted. For each
submenu, three to five selections will appear across the bottom
of the display screen. Many submenus include the File selec-
tion.
Step 3: Press
below the selection titled FILE. Then,
continue to scroll through the list of stored data files until the
desired file is selected.
Step 4: The active file will be highlighted in the File Selection
Window.
5.1.2
Interpreting the DR Mode / RECORD Submenu
Display
When you open a data file, in the DR Mode, the display
should look something like the one shown in Figure 5-3. The
display screen shown in Figure 5-3 is typical of the display
for a Boiler type file, if the RECORD Submenu is activated.
The shaded portion of the display will change based on the
type of data file that’s open. If the open data file is not a
Boiler type, the shaded area of display shown in Figure 5-3
will be replaced with one of the displays shown in Figure 5-4.
Compare the display on your DMS 2 with the display’s
shown in Figures 5-3 and 5-4. Note that the ITEMS listed in
the shaded portion of the display differ from one file-type to
another, and that your display will look different if the RECORD
Submenu is not selected. Also, there are two Navigation
Window display types available in the DR Mode. These
options, SPREAD and LIST VIEWS, are accessible through
the RECORD Submenu and are explained in Section 5.2.6.
Содержание MFE DMS 2
Страница 2: ...DMS 2 DMS 2E Operating Manual GE Inspection Technologies Ultrasonics Ident Nr 021 002 151 Rev H ...
Страница 3: ...THIS PAGE INTENTIONALLY BLANK ...
Страница 22: ...1 Getting Started DMS 2 Operating Manual Page 11 FIGURE 1 7 The VIRTUAL KEYBOARD is shown here ...
Страница 94: ...3 Setting Up the DMS 2 DMS 2 Operating Manual Page 83 FIGURE 3 11 An Auto Labeler screen is shown here ...