5. Working with Stored Data
DMS 2 Operating Manual
Page 155
Step 5: The DR display option will be set to the last option
selected.
NOTE:
Regardless of the instrument settings, all micro-
grids will be displayed in SPREAD VIEW.
5.2.7
Storing and Recalling A-scans
Attaching A-scans to Data Points —The SEND Button’s
Test-Mode Function
Pressing
normally results in a thickness measure-
ment being stored in the active file position. It is also pos-
sible to store an A-scan in the active file position, together
with the thickness measurement, by adjusting the value of
the SEND FUNCTION feature located in the RECORD
Submenu. (Note that there are additional send functions,
related to communication with external-devices, which can
be selected as explained in Chapter 6.)
Step 1: Open the Data Recorder by pushing
.
Step 2: Make sure that the data file you wish to work with is
open.
Step 3: Activate the RECORD Submenu by selecting it with
or
so that it’s highlighted. Selections will appear
across the bottom of the display screen including ITEM and
VALUE.
Step 4: Press
below the selection titled ITEM. You’ll
note that several file characteristics are listed including
SEND FUNCTION.
Step 5: Press
below the selection titled ITEM until
SEND FUNCTION is highlighted.
Step 6: Press
below the selection titled VALUE. Two
options are available:
THICKNESS ONLY – sends only the thickness reading to the
file position when
is pressed.
THCK / THCK+ASCAN – sends the thickness reading and
A-scan to the file position when
is pressed and held
Содержание MFE DMS 2
Страница 2: ...DMS 2 DMS 2E Operating Manual GE Inspection Technologies Ultrasonics Ident Nr 021 002 151 Rev H ...
Страница 3: ...THIS PAGE INTENTIONALLY BLANK ...
Страница 22: ...1 Getting Started DMS 2 Operating Manual Page 11 FIGURE 1 7 The VIRTUAL KEYBOARD is shown here ...
Страница 94: ...3 Setting Up the DMS 2 DMS 2 Operating Manual Page 83 FIGURE 3 11 An Auto Labeler screen is shown here ...