4. Measuring Thickness
DMS 2 Operating Manual
Page 101
Step 1: Press
to activate the FILE NAVIGATION mode.
Step 2: Using the center up/down arrows, or left right arrow
keys, position the bold cell over the thickness reading to be
cleared out.
Step 3: Press
to return to the TEST MENU.
Step 4: Press and hold
for three seconds. The DMS 2
will beep and the GREEN LED will light indicating the
reading has been cleared. The reading number will be
replace with the word EMPTY.
Step 5: Re-take the thickness reading and press
to
store the new thickness value.
If the operator has backed up more than one location, it will
be necessary to use the FILE NAVIGATION mode to get the
active DR location cursor to the next EMPTY location.
Overwriting One or More Readings in a DR File
By turning the OVR selection ON, and then navigating to any
DR location, existing thickness values can be overwritten
with new values without first deleting the existing values. To
overwrite existing thickness values:
Step 1: Access the I/O Submenu located in the TG Second-
ary Menu.
FIGURE 4-5—This is how the DMS 2’s display screen will
appear when a data file is open, SPREAD view is selected,
and the instrument is in the File Navigation Mode. Note that
File Navigation allows you to manually move throughout the
data file, making any point active.
Содержание MFE DMS 2
Страница 2: ...DMS 2 DMS 2E Operating Manual GE Inspection Technologies Ultrasonics Ident Nr 021 002 151 Rev H ...
Страница 3: ...THIS PAGE INTENTIONALLY BLANK ...
Страница 22: ...1 Getting Started DMS 2 Operating Manual Page 11 FIGURE 1 7 The VIRTUAL KEYBOARD is shown here ...
Страница 94: ...3 Setting Up the DMS 2 DMS 2 Operating Manual Page 83 FIGURE 3 11 An Auto Labeler screen is shown here ...