4. Measuring Thickness
Page 102
DMS 2 Operating Manual
Step 2: Set the value of the OVR selection to ON.
Step 3: Activate the FILE NAVIGATION mode and select the
first thickness value you wish to overwrite.
Step 4: Save the new thickness value by pressing
. It
will automatically replace the existing value.
Step 5: You may continue replacing existing thickness
values in this manner.
4.2.3
Closing a Data File
Use this procedure to close an existing data file. Note that a
file can only be opened or closed from the Data Recorder
Menu.
Step 1: Open the Data Recorder Menu accessed from the
Test or TG Mode by pushing
two times. Once the DR
Menu is opened (Figure 4-4), you’ll note five submenus listed
on the Submenu Bar and between three and five selections
listed across the bottom of the display.
Step 2: Activate any submenu located in the DR Menu by
selecting it with
or
so that it’s highlighted. Between
three and five selections will appear across the bottom of the
display screen, including the FILE selection.
Step 3: Press
below the selection titled FILE. In the File-
listing box at the top of the display you’ll note the file option
NO ACTIVE DATA FILE.
Step 4: There will be no active file when NO ACTIVE DATA
FILE is selected.
4.3
Data Enhancement Features
This section describes features that improve the flexibility
and expand the capabilities of the DMS 2. These enhanced
data collection features include the ability to attach com-
ments to a data point, collect additional thickness measure-
ments in a specific location (with the Micro-Grid) option, and
expand the capacity of an established data file.
4.3.1 Attaching Comments to a Data Point
An
Active Comment List (which applies only to the data file
that it’s created for) is a collection of notes that can be
attached to any data point in the file. Before creating a new
file and collecting data points, you should build a Master
Comment List in the DMS 2. A
Master Comment List
contains notes that are used to build an Active Comment
List for any active data file. (Refer to Section 3.4.2 if you
haven’t yet created a Master Comment List.) Once a Master
Comment List has been created, follow these instructions to
Содержание MFE DMS 2
Страница 2: ...DMS 2 DMS 2E Operating Manual GE Inspection Technologies Ultrasonics Ident Nr 021 002 151 Rev H ...
Страница 3: ...THIS PAGE INTENTIONALLY BLANK ...
Страница 22: ...1 Getting Started DMS 2 Operating Manual Page 11 FIGURE 1 7 The VIRTUAL KEYBOARD is shown here ...
Страница 94: ...3 Setting Up the DMS 2 DMS 2 Operating Manual Page 83 FIGURE 3 11 An Auto Labeler screen is shown here ...