GE H
EALTHCARE
D
IRECTION
5394152,
Revision 5
LOGIQ™ P6/P6 P
RO
S
ERVICE
M
ANUAL
Page 8-76
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-31
SOM Replacement
8-2-31-1
Tools
•
Common Phillips screwdrivers
8-2-31-2
Preparations
•
Shut down the system and switch off the main breaker.
8-2-31-3
Removal procedure
1.) Remove the Side Right Cover. Refer to the
8-2-14 "Right or Left Side Cover" on page 8-48
.
2.) Remove the EMI Cover R. Refer to the
8-2-24 "EMI Cover L and R" on page 8-63
3.) Remove the SYSCOMML Assy. Refer to the
8-2-30 "SYSCONML Assy" on page 8-73
4.) Unscrews 4 screws.
5.) Remove the SOM Assy.
Figure 8-116 4 screws for SOM on SYSCON
Figure 8-117 Remove a SOM from SYSCONML Assy