Chapter 10 -
Page 10-23
10-6-7
Isolated Patient Lead (Source) Leakage–Lead to Lead
reference the procedure in the IEC 60601-1. When using the Dale 600, switch the meter’s function
selector to the LEAD-LEAD position. Select and test each of the five ECG lead positions (except ALL)
on the LEAD selector, testing each to the power condition combinations found in the table. Record the
highest leakage current measured.
10-6-7-1
Dale 600 Patient Lead Tests
NEUTRAL POLARITY
1.) Closed Normal
2.) Open Normal
3.) Closed Reversed
4.) Open Reversed