Page 8-91
8-2-42
ARP II Assy 110V/220V
8-2-42-1
Tools
•
Common Phillips screwdrivers
8-2-42-2
Preparations
•
Shut down the system and switch off the main breaker.
8-2-42-3
Removal procedure
1.) Remove side right cover and EMI cover. Refer to the
8-2-14 "Right or Left Side Cover" on page 8-
and
8-2-24 "EMI Cover L and R" on page 8-63
2.) Remove the Rear Cover. Refer to the
8-2-16 "Rear Cover" on page 8-52
.
3.) Unscrew 4 screws (1-4).
4.) Disconnect J15 connector.
Figure 8-133 Screws of the rear panel
Figure 8-134 AEXP Cable of rear panel
(1)
(3)
(4)