GE H
EALTHCARE
D
IRECTION
5394152,
Revision 5
LOGIQ™ P6/P6 P
RO
S
ERVICE
M
ANUAL
Chapter 4 -
Page 4-49
4-9-10
Nest fan Function Validation Procedure
1.) After replace Nest fan assy, turn on the system. Refer to the
3-5-1 "Power On/Boot Up" on page 3-
.
2.) Confirm the fan rotating with naked eye and check the fan operating by rotating sound. No strange
sound should be heard.
NOTE:
Do not close the right EMI bracket and right plastic cover before check the fan operation.
4-9-11
Transbox/ Fan Function Validation Procedure
1.) After replace Transbox fan assy, turn on the system. Refer to the
2.) Confirm the fan rotating by feeling wind. Refer to the figure below.
NOTE:
Remove right side dusty filter and check the fan operation.
Figure 4-47 Checking Nest fan rotating
Figure 4-48 Checking Transbox fan rotating