GE H
EALTHCARE
D
IRECTION
5394152,
Revision 5
LOGIQ™ P6/P6 P
RO
S
ERVICE
M
ANUAL
Chapter 2 -
Page 2-3
2-2-2
Electrical Requirements
NOTE:
GE Healthcare requires a dedicated power and ground for the proper operation of its Ultrasound
equipment. This dedicated power shall originate at the last distribution panel before the system.
Sites with a mains power system with defined Neutral and Live:
The dedicated line shall consist of one phase, a neutral (not shared with any other circuit), and a full
size ground wire from the distribution panel to the Ultrasound outlet.
Sites with a mains power system without a defined Neutral:
The dedicated line shall consist of one phase (two lines), not shared with any other circuit, and a full
size ground wire from the distribution panel to the Ultrasound outlet.
NOTE:
Please note that image artifacts can occur, if at any time within the facility, the ground from the
main facility's incoming power source to the Ultrasound unit is only a conduit.
2-2-2-1
LOGIQ™ P6/P6 Pro Power Requirements
The following power line parameters should be monitored for one week before installation. We
recommend that you use an analyzer Dranetz Model 606-3 or Dranetz Model 626:
2-2-2-2
Inrush Current
Inrush Current is not a factor to consider due to the inrush current limiting properties of the power
supplies.
2-2-2-3
Site Circuit Breaker
It is recommended that the branch circuit breaker for the machine be ready accessible.
Table 2-4
Electrical Specifications for LOGIQ™ P6/P6 Pro
PARAMETER
AREA
LIMITS
Voltage Range
100~120V
100~120 VAC
±
10% (90-132 VAC)
220~240V
220-240 VAC
±
10% (198-264 VAC)
Power
All applications
MAX. 950 VA
Line Frequency
All applications
50/60Hz (
±
2Hz)
Power Transients
All applications
Less than 25% of nominal peak voltage
for less than 1 millisecond for any type of
transient, including line frequency,
synchronous, asynchronous, or
aperiodic transients.
Decaying Oscillation
All applications
Less than 15% of peak voltage for less
than 1 millisecond.
CAUTION
POWER OUTAGE MAY OCCURE.
The LOGIQ™ P6/P6 Pro requires a dedicated single branch circuit. To avoid circuit overload and
possible loss of critical care equipment, make sure you DO NOT have any other equipment
operating on the same circuit.