Page 5-40
Section 5-1 - Overview
Section 5-9Machine RFS
5-9-1
General
This describes general MRFS (Machine RFS) functionality.
1.) MRFS function is an automatic call for service from the system to the OLC when system detects
high temperature or high or low voltage.
5-9-2
Enabling MRFS
NOTE:
In order to enable MRFS, System must be connected to network and should be checked out to
a product server. Default MRFS user should also be set.
1.) If a system is not already checked out, the system should be checked out to a product server. Refer
to the
5-8-2 - System Checkout Procedure.
2.) MRFS function is an automatic call for service from the system to the OLC when system detects
high temperature or high or low voltage.
3.) Right Click on “Connect to GE” icon.
4.) When RFS window pops up, click on “Users” tab.
5.) Press “Add User”.
Figure 5-42 Contact GE Icon
Figure 5-43 User tab
Figure 5-44 Add User