Chapter 8 -
Page 8-43
8-2-10
Freeze key assy
8-2-10-1
Tools
•
Common Phillips screwdrivers
8-2-10-2
Preparations
•
Shut down the system and switch off the main breaker.
8-2-10-3
Removal procedure
1.) Remove the Main Keyboard Assy. Refer to the
8-2-6 "Main Keyboard Assy" on page 8-37
2.) Remove the 3 screws (1-3) and disconnect 1 cable.
3.) Perform the following functional tests. If all are successful, include the debrief script provided below.
8-2-10-4
Mounting Procedure
Install the new parts in the reverse order of removal.
Figure 8-79 Freeze Key in the main keyboard
Figure 8-80 Freeze Key connector
Table 8-11 Functional Tests
Service Manual
Section
Functional Test / Diagnostic Test
Debrief Script
“Service Manual, Direction
5394152, Rev 1+, Section 8-2-10. Equipment
passed all required tests and is ready for use. “