Appendix C: Operating Theory
205
OPERATING THEORY
THEORY
The iVent
TM
201 is a positive pressure mechanical ventilator. This appendix
describes the ventilator’s components and depicts a schematic flow of the
electro-pneumatic system. The function of each component is described below.
AMBIENT AIR FILTER
Air is drawn from the external environment through this filter. The filter captures
particles greater than 5 microns with an efficiency of 99.0%.
CBRN FILTER (OPTIONAL)
An optional configuration enables air to be drawn in through a Chemical,
Biological, Radiological, and Nuclear (CBRN) filter.
LOW PRESSURE O2 ADAPTER AND FILTER
(OPTIONAL)
An optional configuration that enables the usage of low pressure O2 sources,
such as concentrators, to provide O2 enrichment.
AIR/OXYGEN BLENDING SYSTEM
The blending system consists of an air entrainment port and a PSV (Proportional
Solenoid Valve) High Pressure O2 Enrichment Valve. The ventilator software
controls both the port and PSV account for variations in the inspired flow to
achieve the required oxygen concentration.
NOTE
Units with serial numbers below 14999 have a different type of Oxygen
blender based on a demand valve and slider, drive by a stepper motor.
INLET MANIFOLD
The inlet manifold houses the air entrainment port, PSV or demand valve, and
fail-safe bypass port.
Содержание iVent 201 Series
Страница 1: ...GE Healthcare iVentTM201 User s Reference Manual Revision 13...
Страница 2: ...VersaMed a General Electric Company doing business as GE Healthcare...
Страница 4: ......
Страница 14: ......
Страница 42: ......
Страница 68: ...56 Notes...
Страница 70: ......
Страница 98: ...86 Notes...
Страница 100: ......
Страница 140: ...128 Notes...
Страница 142: ......
Страница 152: ......
Страница 154: ......
Страница 184: ......
Страница 201: ...7 Care Maintenance and Tests 189 Notes...
Страница 202: ......
Страница 203: ...Appendix A Glossary 191 APPENDIX A GLOSSARY...
Страница 204: ......
Страница 210: ...198 Notes...
Страница 211: ...Appendix B Warranty 199 APPENDIX B WARRANTY...
Страница 212: ......
Страница 215: ...Appendix C Operating Theory 203 APPENDIX C OPERATING THEORY...
Страница 216: ......
Страница 219: ...Appendix C Operating Theory 207...
Страница 220: ......
Страница 221: ...Appendix C Operating Theory 209 Notes...
Страница 222: ......
Страница 223: ...Appendix D Theory of Breath Delivery 211 APPENDIX D THEORY OF BREATH DELIVERY...
Страница 224: ......
Страница 233: ...Appendix D Theory of Breath Delivery 221 Notes...
Страница 234: ......
Страница 235: ...Appendix E Ventilation Modes 223 APPENDIX E VENTILATION MODES...
Страница 236: ......
Страница 243: ...Appendix E Ventilation Modes 231 Notes...
Страница 244: ......
Страница 246: ......
Страница 252: ...240 Notes...
Страница 253: ...Appendix G Accessories 241 APPENDIX G ACCESSORIES...
Страница 254: ......
Страница 276: ...264 5 Fasten the external battery using the strap belt Figure 199 Connecting the Optional Battery Bag...
Страница 277: ...Appendix H Remote Alarm Connector 265 APPENDIX H REMOTE ALARM CONNECTOR...
Страница 278: ......
Страница 280: ......
Страница 281: ...Appendix I Part Numbers and Accessories 269 APPENDIX I PART NUMBERS AND ACCESSORIES...
Страница 282: ......
Страница 286: ...274...
Страница 289: ...Index 277...