60
NOTE
Parameters for all the
iVent
TM
201
’s possible ventilation modes are set when
you choose a patient weight upon startup. For a view of the default
parameters according to weight, refer to Table 1: Default Start Parameters for
Volume Ventilation for 10 kg to 70 kg through Table 4, page 107.
The default selection for patient weight range for the
iVent
TM
201
is 70+kg. This
may be changed using the
Configuration
utility described on page 185.
As a safety measure, within a few seconds after the patient weight selection pop-
up window appears, the iVent
TM
201 generates airflow to see if resistance is
encountered in the patient circuit, which means that the patient is connected.
NOTE
This is an optional feature, not available in all
iVent
TM
201
units.
If the iVent
TM
201 detects resistance, it sounds an alarm and begins
ventilating in SIMV pressure control mode at a rate of 15 breaths per minute,
P(INSP) 20 and a PEEP of 5.
If the iVent
TM
201 detects no airway resistance, it enters
Standby
mode.
During this state, if flow is detected in the patient circuit the iVent
TM
201
performs an auto-start.
O.V.T
The patient circuit must be tested each time it is connected, so if you are
reconnecting a patient circuit or using a new one, you must perform the
Operational Verification Test (O.V.T.). This test, which takes less than one minute,
checks the integrity of the breathing circuit and audible alarm functionality.
Two plastic caps for covering the ends of the patient circuit are required for the
O.V.T. They are included with all patient circuits offered by Versamed.
To perform an O.V.T.:
1.
Rotate the Control Knob until the
O.V.T
. option is selected in bright blue.
2.
Press the Control Knob until you hear a click. The
O.V.T. Instructions
pop up
window appears (Figure 29).
Содержание iVent 201 Series
Страница 1: ...GE Healthcare iVentTM201 User s Reference Manual Revision 13...
Страница 2: ...VersaMed a General Electric Company doing business as GE Healthcare...
Страница 4: ......
Страница 14: ......
Страница 42: ......
Страница 68: ...56 Notes...
Страница 70: ......
Страница 98: ...86 Notes...
Страница 100: ......
Страница 140: ...128 Notes...
Страница 142: ......
Страница 152: ......
Страница 154: ......
Страница 184: ......
Страница 201: ...7 Care Maintenance and Tests 189 Notes...
Страница 202: ......
Страница 203: ...Appendix A Glossary 191 APPENDIX A GLOSSARY...
Страница 204: ......
Страница 210: ...198 Notes...
Страница 211: ...Appendix B Warranty 199 APPENDIX B WARRANTY...
Страница 212: ......
Страница 215: ...Appendix C Operating Theory 203 APPENDIX C OPERATING THEORY...
Страница 216: ......
Страница 219: ...Appendix C Operating Theory 207...
Страница 220: ......
Страница 221: ...Appendix C Operating Theory 209 Notes...
Страница 222: ......
Страница 223: ...Appendix D Theory of Breath Delivery 211 APPENDIX D THEORY OF BREATH DELIVERY...
Страница 224: ......
Страница 233: ...Appendix D Theory of Breath Delivery 221 Notes...
Страница 234: ......
Страница 235: ...Appendix E Ventilation Modes 223 APPENDIX E VENTILATION MODES...
Страница 236: ......
Страница 243: ...Appendix E Ventilation Modes 231 Notes...
Страница 244: ......
Страница 246: ......
Страница 252: ...240 Notes...
Страница 253: ...Appendix G Accessories 241 APPENDIX G ACCESSORIES...
Страница 254: ......
Страница 276: ...264 5 Fasten the external battery using the strap belt Figure 199 Connecting the Optional Battery Bag...
Страница 277: ...Appendix H Remote Alarm Connector 265 APPENDIX H REMOTE ALARM CONNECTOR...
Страница 278: ......
Страница 280: ......
Страница 281: ...Appendix I Part Numbers and Accessories 269 APPENDIX I PART NUMBERS AND ACCESSORIES...
Страница 282: ......
Страница 286: ...274...
Страница 289: ...Index 277...