6 Alarms
163
Warning
Set Count or
Time of
Condition
Mode of Failure Priority Probable Cause Criterion for
Clearing Alarm
Corrective Action
Low Pressure
3 breath delay
(4 with setting
change)
Airway pressure
during inspiration
is lower than the
Low Pressure
alarm threshold
Mid
Clinical condition
Incorrect alarm
setting
Patient
disconnect or
high leak
Pressure is
higher than
alarm threshold
for 2 breaths
Subject to clinical
judgment, correct
setting as
necessary
Verify patient
circuit
Low SpO2
2 breaths
delay
SpO2 level is
below the set
lower SpO2 limit
High
The patient
clinical condition,
FiO
2
level set too
low, or SpO2
alarm setting too
high
SpO2 level is
higher than the
alarm limit from
1 breath.
Assessing the
patient clinical
condition,
increasing the
FiO
2
concentration, or
decreasing the
SpO2 alarm
setting.
Auto Start
Patient is
connected but
the unit is in
Standby mode
High
Mid
User enters
Standby mode
unintentionally
User connects
the ventilator to
the patient but
fails to press
Start
N/A Clear
the
alarm
and set the
required
ventilation
parameters
Service Notice Technical
irregularity
(needs Technical
Support Services)
High Power
surge
Multiple sensor
disconnection
Calibration
required
Ventilator
Malfunction
N/A
Turn the system
OFF and ON
again If the alarm
is not cleared,
remove the unit
for inspection and
servicing by
qualified
technical
personnel
Warm start-up
service notice
Software
detected
technical
irregularity
High Software
exception
N/A 1.
Single event –
no action
required –
resumed
ventilation with
previous
settings in up
to 30 seconds.
2.
Multiple
repeated event
– remove the
unit for
inspection and
servicing by
Содержание iVent 201 Series
Страница 1: ...GE Healthcare iVentTM201 User s Reference Manual Revision 13...
Страница 2: ...VersaMed a General Electric Company doing business as GE Healthcare...
Страница 4: ......
Страница 14: ......
Страница 42: ......
Страница 68: ...56 Notes...
Страница 70: ......
Страница 98: ...86 Notes...
Страница 100: ......
Страница 140: ...128 Notes...
Страница 142: ......
Страница 152: ......
Страница 154: ......
Страница 184: ......
Страница 201: ...7 Care Maintenance and Tests 189 Notes...
Страница 202: ......
Страница 203: ...Appendix A Glossary 191 APPENDIX A GLOSSARY...
Страница 204: ......
Страница 210: ...198 Notes...
Страница 211: ...Appendix B Warranty 199 APPENDIX B WARRANTY...
Страница 212: ......
Страница 215: ...Appendix C Operating Theory 203 APPENDIX C OPERATING THEORY...
Страница 216: ......
Страница 219: ...Appendix C Operating Theory 207...
Страница 220: ......
Страница 221: ...Appendix C Operating Theory 209 Notes...
Страница 222: ......
Страница 223: ...Appendix D Theory of Breath Delivery 211 APPENDIX D THEORY OF BREATH DELIVERY...
Страница 224: ......
Страница 233: ...Appendix D Theory of Breath Delivery 221 Notes...
Страница 234: ......
Страница 235: ...Appendix E Ventilation Modes 223 APPENDIX E VENTILATION MODES...
Страница 236: ......
Страница 243: ...Appendix E Ventilation Modes 231 Notes...
Страница 244: ......
Страница 246: ......
Страница 252: ...240 Notes...
Страница 253: ...Appendix G Accessories 241 APPENDIX G ACCESSORIES...
Страница 254: ......
Страница 276: ...264 5 Fasten the external battery using the strap belt Figure 199 Connecting the Optional Battery Bag...
Страница 277: ...Appendix H Remote Alarm Connector 265 APPENDIX H REMOTE ALARM CONNECTOR...
Страница 278: ......
Страница 280: ......
Страница 281: ...Appendix I Part Numbers and Accessories 269 APPENDIX I PART NUMBERS AND ACCESSORIES...
Страница 282: ......
Страница 286: ...274...
Страница 289: ...Index 277...