Figure 165: Pressure and Flow Triggering by Patient
1.
Pressure trigger
. The breath is initiated when airway pressure drops below
the baseline in an amount greater than the set sensitivity value. (a in the
figure above).
2.
Flow trigger
. The breath is initiated when initial inspiratory flow generated by
patient’s effort is greater than the set sensitivity value. (b in the figure
above).
3.
Dual trigger
. The breath is initiated when either flow or pressure exceeds the
set sensitivity values.
The triggering type and level of effort required to initiate a breath are set using
the Sensitivity Control. When the dual pressure/flow trigger is selected, the
patient breath can be initiated by flow or pressure sensitivity levels which ever is
exceeded first.
For patient safety it is not possible to disable flow or pressure triggering when
using the Adaptive Bi-level Mode.
BREATH TYPES
The ventilator delivers two primary breath types:
1.
Mandatory Breath. A breath in which the ventilator controls and terminates
phase. This breath type is divided into three secondary types according to the
method by which the inspiratory phase is initiated:
Ventilator breath
- initiated by the ventilator (time triggered). A blue fan icon
appears in the VT display.
Assist breath
- initiated by the patient (pressure or flow triggered). A pink fan
icon appears in the VT display.
Manual breath
- initiated by the operator by pressing the manual breath
button. A blue fan icon appears in the VT display.
Each of the above breaths may be delivered in one of two inspiratory methods:
Volume control breath
- mandatory breath in which the ventilator delivers
the tidal volume set during a certain time interval at a set or Adaptive
inspiratory flow.
Содержание iVent 201 Series
Страница 1: ...GE Healthcare iVentTM201 User s Reference Manual Revision 13...
Страница 2: ...VersaMed a General Electric Company doing business as GE Healthcare...
Страница 4: ......
Страница 14: ......
Страница 42: ......
Страница 68: ...56 Notes...
Страница 70: ......
Страница 98: ...86 Notes...
Страница 100: ......
Страница 140: ...128 Notes...
Страница 142: ......
Страница 152: ......
Страница 154: ......
Страница 184: ......
Страница 201: ...7 Care Maintenance and Tests 189 Notes...
Страница 202: ......
Страница 203: ...Appendix A Glossary 191 APPENDIX A GLOSSARY...
Страница 204: ......
Страница 210: ...198 Notes...
Страница 211: ...Appendix B Warranty 199 APPENDIX B WARRANTY...
Страница 212: ......
Страница 215: ...Appendix C Operating Theory 203 APPENDIX C OPERATING THEORY...
Страница 216: ......
Страница 219: ...Appendix C Operating Theory 207...
Страница 220: ......
Страница 221: ...Appendix C Operating Theory 209 Notes...
Страница 222: ......
Страница 223: ...Appendix D Theory of Breath Delivery 211 APPENDIX D THEORY OF BREATH DELIVERY...
Страница 224: ......
Страница 233: ...Appendix D Theory of Breath Delivery 221 Notes...
Страница 234: ......
Страница 235: ...Appendix E Ventilation Modes 223 APPENDIX E VENTILATION MODES...
Страница 236: ......
Страница 243: ...Appendix E Ventilation Modes 231 Notes...
Страница 244: ......
Страница 246: ......
Страница 252: ...240 Notes...
Страница 253: ...Appendix G Accessories 241 APPENDIX G ACCESSORIES...
Страница 254: ......
Страница 276: ...264 5 Fasten the external battery using the strap belt Figure 199 Connecting the Optional Battery Bag...
Страница 277: ...Appendix H Remote Alarm Connector 265 APPENDIX H REMOTE ALARM CONNECTOR...
Страница 278: ......
Страница 280: ......
Страница 281: ...Appendix I Part Numbers and Accessories 269 APPENDIX I PART NUMBERS AND ACCESSORIES...
Страница 282: ......
Страница 286: ...274...
Страница 289: ...Index 277...