148
Figure 135: The Minute Volume Alarm Slider Selected
4.
Press the Control Knob to change the alarm settings for the
Minute Volume
.
5.
The top of the slider is selected bright blue. Turn the dial to adjust the alarm
for the upper limit in liters per minute. Notice if you attempt to adjust the
upper limit of the alarm below the currently-selected lower limit, you push
the lower limit down. Setting the
Minute Volume
alarm over the maximum
level of 99 turns off the alarm’s upper limit.
6.
Press the Control Knob to confirm your selection.
7.
The bottom of the slider is selected bright blue. Turn the dial to adjust the
alarm for the lower limit in liters per minute. Note that adjusting the lower
limit higher than the currently-selected higher limit moves the higher limit
up.
8.
Press the Control Knob to confirm the settings.
9.
The
Alarm Setting
screen is re-displayed. You can dial and select an
additional alarm setting to adjust, or dial and select
Accept
to accept all the
alarm settings and return to the
Main
screen.
Setting the Pressure Alarm
To set the pressure alarm upper and lower limits:
1.
Select and press the
Menu
option in the
Main
screen to view the
Main Menu
.
2.
Select and press the first item on the
Main Menu
,
Alarm Settings
.
3.
In the
Alarm Settings
screen, dial the Control Knob until
Pressure
is selected
in relief (Figure 136).
Содержание iVent 201 Series
Страница 1: ...GE Healthcare iVentTM201 User s Reference Manual Revision 13...
Страница 2: ...VersaMed a General Electric Company doing business as GE Healthcare...
Страница 4: ......
Страница 14: ......
Страница 42: ......
Страница 68: ...56 Notes...
Страница 70: ......
Страница 98: ...86 Notes...
Страница 100: ......
Страница 140: ...128 Notes...
Страница 142: ......
Страница 152: ......
Страница 154: ......
Страница 184: ......
Страница 201: ...7 Care Maintenance and Tests 189 Notes...
Страница 202: ......
Страница 203: ...Appendix A Glossary 191 APPENDIX A GLOSSARY...
Страница 204: ......
Страница 210: ...198 Notes...
Страница 211: ...Appendix B Warranty 199 APPENDIX B WARRANTY...
Страница 212: ......
Страница 215: ...Appendix C Operating Theory 203 APPENDIX C OPERATING THEORY...
Страница 216: ......
Страница 219: ...Appendix C Operating Theory 207...
Страница 220: ......
Страница 221: ...Appendix C Operating Theory 209 Notes...
Страница 222: ......
Страница 223: ...Appendix D Theory of Breath Delivery 211 APPENDIX D THEORY OF BREATH DELIVERY...
Страница 224: ......
Страница 233: ...Appendix D Theory of Breath Delivery 221 Notes...
Страница 234: ......
Страница 235: ...Appendix E Ventilation Modes 223 APPENDIX E VENTILATION MODES...
Страница 236: ......
Страница 243: ...Appendix E Ventilation Modes 231 Notes...
Страница 244: ......
Страница 246: ......
Страница 252: ...240 Notes...
Страница 253: ...Appendix G Accessories 241 APPENDIX G ACCESSORIES...
Страница 254: ......
Страница 276: ...264 5 Fasten the external battery using the strap belt Figure 199 Connecting the Optional Battery Bag...
Страница 277: ...Appendix H Remote Alarm Connector 265 APPENDIX H REMOTE ALARM CONNECTOR...
Страница 278: ......
Страница 280: ......
Страница 281: ...Appendix I Part Numbers and Accessories 269 APPENDIX I PART NUMBERS AND ACCESSORIES...
Страница 282: ......
Страница 286: ...274...
Страница 289: ...Index 277...