160
Warning
Set Count or
Time of
Condition
Mode of Failure Priority Probable Cause Criterion for
Clearing Alarm
Corrective Action
High O
2
20 breath
delay
Oxygen
concentration
exceeds the high
alarm setting
High
Improper alarm
setting
Change in
patient
breathing
pattern with low
flow adapter
O
2
sensor out of
calibration
FiO
2
is lower
than alarm
setting for 2
breaths
Adjust alarm
setting
Verify O
2
concentration
with external
analyzer
Refer unit to
qualified service
technician for
calibration of O
2
sensor
High Heart
Rate
2 breath delay
The heart rate
exceeds the
upper rate limit
Mid
Clinical Condition Heart rate is
lower than the
alarm settings
for 1 breath
Check the
patient’s
condition.
High Minute
Volume
4 breath delay
(20
with setting
change)
Minute volume
exceeds high
alarm setting
High Clinical
condition
Alarm set too low
Breath rate set
too high
Volume is lower
than alarm
setting, for
more than 2
breaths
Assess patient
Increase alarm
setting if
appropriate
Decrease trigger
sensitivity
Содержание iVent 201 Series
Страница 1: ...GE Healthcare iVentTM201 User s Reference Manual Revision 13...
Страница 2: ...VersaMed a General Electric Company doing business as GE Healthcare...
Страница 4: ......
Страница 14: ......
Страница 42: ......
Страница 68: ...56 Notes...
Страница 70: ......
Страница 98: ...86 Notes...
Страница 100: ......
Страница 140: ...128 Notes...
Страница 142: ......
Страница 152: ......
Страница 154: ......
Страница 184: ......
Страница 201: ...7 Care Maintenance and Tests 189 Notes...
Страница 202: ......
Страница 203: ...Appendix A Glossary 191 APPENDIX A GLOSSARY...
Страница 204: ......
Страница 210: ...198 Notes...
Страница 211: ...Appendix B Warranty 199 APPENDIX B WARRANTY...
Страница 212: ......
Страница 215: ...Appendix C Operating Theory 203 APPENDIX C OPERATING THEORY...
Страница 216: ......
Страница 219: ...Appendix C Operating Theory 207...
Страница 220: ......
Страница 221: ...Appendix C Operating Theory 209 Notes...
Страница 222: ......
Страница 223: ...Appendix D Theory of Breath Delivery 211 APPENDIX D THEORY OF BREATH DELIVERY...
Страница 224: ......
Страница 233: ...Appendix D Theory of Breath Delivery 221 Notes...
Страница 234: ......
Страница 235: ...Appendix E Ventilation Modes 223 APPENDIX E VENTILATION MODES...
Страница 236: ......
Страница 243: ...Appendix E Ventilation Modes 231 Notes...
Страница 244: ......
Страница 246: ......
Страница 252: ...240 Notes...
Страница 253: ...Appendix G Accessories 241 APPENDIX G ACCESSORIES...
Страница 254: ......
Страница 276: ...264 5 Fasten the external battery using the strap belt Figure 199 Connecting the Optional Battery Bag...
Страница 277: ...Appendix H Remote Alarm Connector 265 APPENDIX H REMOTE ALARM CONNECTOR...
Страница 278: ......
Страница 280: ......
Страница 281: ...Appendix I Part Numbers and Accessories 269 APPENDIX I PART NUMBERS AND ACCESSORIES...
Страница 282: ......
Страница 286: ...274...
Страница 289: ...Index 277...