7 Care, Maintenance and Tests
173
INTRODUCTION
This section shows you how to:
Clean and maintain the iVent
TM
201
Keep the iVent
TM
201 operable with the help of planned maintenance
Calibrate and perform diagnostic tests on the iVent
TM
201
CLEANING AND MAINTENANCE
While the iVent
TM
201 has been designed to resist damage, staining, and wear, it is
recommended to perform occasional cleaning and basic maintenance
procedures to enhance its lifecycle.
The following chart summarizes cleaning and routine maintenance procedures
and, where appropriate, time intervals.
Table 12: Maintenance Procedures and Intervals
Part
Procedure
Comments
Ventilator
Wipe exterior clean with
a damp cloth and mild
detergent.
Air Inlet Filter
Replace every 500 hours
(or 1 month) of operation
or as necessary.
Do not attempt to
clean or reuse the air
inlet filter.
Cooling Vents &
Cooling Inlet
Filter
Clean every 1500 hours
(or 3 months) of use, or
as necessary.
Use vacuum device to
clean the vents and
cooling air inlet filter.
Battery Pack
Recharge after every
three months of storage.
Replace every year or as
necessary.
Actual life depends on
history of use and
storage.
O
2
sensor
Calibrate every 6 months It is good practice to
calibrate every 3
months
Replace every 2 years
Perform O2 Calibration
and VVT after unit
returns from service
Pneumatic unit Replace every 15,000
hours of operation or
every 4 years, whichever
comes first.
Inlet and outlet
mufflers
Replace annually or as
necessary
Actual life depends on
history of use and
storage.
Other
accessories
Follow manufacturer instructions.
Содержание iVent 201 Series
Страница 1: ...GE Healthcare iVentTM201 User s Reference Manual Revision 13...
Страница 2: ...VersaMed a General Electric Company doing business as GE Healthcare...
Страница 4: ......
Страница 14: ......
Страница 42: ......
Страница 68: ...56 Notes...
Страница 70: ......
Страница 98: ...86 Notes...
Страница 100: ......
Страница 140: ...128 Notes...
Страница 142: ......
Страница 152: ......
Страница 154: ......
Страница 184: ......
Страница 201: ...7 Care Maintenance and Tests 189 Notes...
Страница 202: ......
Страница 203: ...Appendix A Glossary 191 APPENDIX A GLOSSARY...
Страница 204: ......
Страница 210: ...198 Notes...
Страница 211: ...Appendix B Warranty 199 APPENDIX B WARRANTY...
Страница 212: ......
Страница 215: ...Appendix C Operating Theory 203 APPENDIX C OPERATING THEORY...
Страница 216: ......
Страница 219: ...Appendix C Operating Theory 207...
Страница 220: ......
Страница 221: ...Appendix C Operating Theory 209 Notes...
Страница 222: ......
Страница 223: ...Appendix D Theory of Breath Delivery 211 APPENDIX D THEORY OF BREATH DELIVERY...
Страница 224: ......
Страница 233: ...Appendix D Theory of Breath Delivery 221 Notes...
Страница 234: ......
Страница 235: ...Appendix E Ventilation Modes 223 APPENDIX E VENTILATION MODES...
Страница 236: ......
Страница 243: ...Appendix E Ventilation Modes 231 Notes...
Страница 244: ......
Страница 246: ......
Страница 252: ...240 Notes...
Страница 253: ...Appendix G Accessories 241 APPENDIX G ACCESSORIES...
Страница 254: ......
Страница 276: ...264 5 Fasten the external battery using the strap belt Figure 199 Connecting the Optional Battery Bag...
Страница 277: ...Appendix H Remote Alarm Connector 265 APPENDIX H REMOTE ALARM CONNECTOR...
Страница 278: ......
Страница 280: ......
Страница 281: ...Appendix I Part Numbers and Accessories 269 APPENDIX I PART NUMBERS AND ACCESSORIES...
Страница 282: ......
Страница 286: ...274...
Страница 289: ...Index 277...