4.3.24 “Up Pedal Pressed” Alarm
Cause or Circumstances
Failure
Action
The bed up pedal switch was
pressed and held on during
power up.
None
When the bed up pedal switch is released, the alarm
deactivates. If the alarm silence button is pressed or the
pedal remains on for 100 seconds, the Bed Up Pedal
Failure alarm activates and the Bed Up Pedal Switches
are disabled. Shut down the unit and power up to clear
failure.
Other
Bed up pedal switch has
failed.
Refer to section 4.3.8 for troubleshooting information.
4.4 Error Codes
The following sections describe the meaning of each system failure code for the Giraffe OmniBed. Each table
lists the possible causes and the corresponding recommended actions, in the order in which they should be
tried.
For information on Giraffe Scale error codes, refer to the Giraffe Scale Service Manual, which can be ordered as
part of a Giraffe OmniBed service manual kit: (US) 6600-0834-801 or (INTL) 6600-0834-802.
4.4.1 System Failure 0
System failure 0 indicates: Communication error between the display driver board and the control board.
Cause(s)
Action(s)
Defective scale.
Disconnect the scale connector at probe panel. Power cycle the unit. If the error clears,
replace the scale.
Defective scale
harness.
Disconnect the scale harness from J7 on the control board. (The scale may be plugged in.)
Power cycle the unit. If the error clears, replace the scale harness.
Defective Servo Oxygen
board.
Disconnect the RS485 harness from J8 on the Servo Oxygen board. Power cycle the unit. If the
error clears, replace the Servo Oxygen Board.
Defective Servo Oxygen
cable.
Disconnect RS485 bus cable from J6 on the control board. Power cycle the unit. If the error
clears, replace the RS485 cable going to the Servo Oxygen board.
Defective battery.
The battery is used only for power failure. Try powering up the unit with the battery
disconnected. If the error clears, replace the battery.
Defective 14 pin ribbon
cable between control
board and display
driver board.
Check that the 14 pin ribbon cable is properly seated at J6 on the control board and at J20 on
the display driver board. (Have all three parts on hand and try them one at a time. Try routing
the cable around the outside of unit first, then try the display board, and then try the control
board. If the error is intermittent, replace all three.)
Defective display driver
board.
Replace display driver board.
Defective control
board.
Replace control board.
© 2001 by Datex-Ohmeda, Inc.. All rights reserved.
6600-0343-000 104
43
Chapter 4: Troubleshooting
Содержание Giraffe OmniBed
Страница 10: ...x 6600 0343 000 104 2001 by Datex Ohmeda Inc All rights reserved Table of Contents ...
Страница 28: ...xxviii 6600 0343 000 104 2001 by Datex Ohmeda Inc All rights reserved About this Manual ...
Страница 38: ...10 6600 0343 000 104 2001 by Datex Ohmeda Inc All rights reserved Chapter 1 Functional Description ...
Страница 48: ...20 6600 0343 000 104 2001 by Datex Ohmeda Inc All rights reserved Chapter 2 Service Checkout ...
Страница 56: ...28 6600 0343 000 104 2001 by Datex Ohmeda Inc All rights reserved Chapter 3 Calibration and Maintenance ...
Страница 260: ...232 6600 0343 000 104 2001 by Datex Ohmeda Inc All rights reserved Chapter 6 Illustrated Parts ...