Symptom
Possible Cause
Action
Elevating base will
not go up or down,
canopy works OK.
Pedal switches are disabled on
nurse setup screen or service
screen.
Check Setup screen (wrench symbol) or service screen to see if
pedal switches are disabled.
Defective elevating base
motor.
Check the voltage to the elevating base motor. Refer to the Tips
section (4.6.9). If the voltage is OK, replace the elevating base
assembly.
Defective relay board.
Replace relay board.
Canopy will not go
up or down with
pedal switches or
hand switches,
Elevating base
works OK.
Canopy motor turns. Defective
belt or clutch.
Check rail drive belt and clutch. Refer to the Tips section (4.6.10).
Canopy motor does not turn.
Defective canopy motor
Check the voltage to the canopy motor. Refer to the Tips section
(4.6.9). If voltage is OK, replace the canopy motor.
Canopy motor does not turn.
Defective 50 pin ribbon cable
between the relay and control
board.
Check for continuity on pins 23, 24, 26, 27, 34 (motor control and
safety relay 2 control signals). If defective, replace the cable.
Canopy motor does not turn.
Defective relay board.
Replace relay board.
Canopy works in
one direction but
not the other.
Defective canopy motor.
Check the voltage to the canopy motor for both up and down
movement. Refer to the Tips section (4.6.9). If OK, replace the
canopy motor/gearbox.
Defective 50 pin ribbon cable
between the relay and control
board.
Check for continuity on pins 23, 24, 26, 27 (motor control signals).
If defective, replace the cable.
Defective relay board.
Replace relay board.
Canopy travels
down part way then
stops or goes very
slow for the last
part.
An object fell into the opening
behind the radiant heater
floppy door.
Check that the floppy door, located behind the radiant heater
moves freely. Look to be sure nothing has fallen behind the door.
Canopy rises slowly
and stops before
doors are opened
fully.
Broken rungs in rail drive belt
Inspect belt. Refer to the Tips section (4.6.10).
Canopy is clouded. Use of alcohol or hand-
disinfectants
Replace canopy.
Unit will not switch
to Baby Control
Mode.
A patient probe is plugged into
Jack 2 on the probe panel.
Disconnect probe from Jack 2 (unit will only allow Baby Control
operation with a single probe in Jack 1).
Baby Control is disabled on
Service screen.
Select BOTH for Patient Control on service screen to activate
Baby Control.
56
6600-0343-000 104
© 2001 by Datex-Ohmeda, Inc.. All rights reserved.
Chapter 4: Troubleshooting
Содержание Giraffe OmniBed
Страница 10: ...x 6600 0343 000 104 2001 by Datex Ohmeda Inc All rights reserved Table of Contents ...
Страница 28: ...xxviii 6600 0343 000 104 2001 by Datex Ohmeda Inc All rights reserved About this Manual ...
Страница 38: ...10 6600 0343 000 104 2001 by Datex Ohmeda Inc All rights reserved Chapter 1 Functional Description ...
Страница 48: ...20 6600 0343 000 104 2001 by Datex Ohmeda Inc All rights reserved Chapter 2 Service Checkout ...
Страница 56: ...28 6600 0343 000 104 2001 by Datex Ohmeda Inc All rights reserved Chapter 3 Calibration and Maintenance ...
Страница 260: ...232 6600 0343 000 104 2001 by Datex Ohmeda Inc All rights reserved Chapter 6 Illustrated Parts ...