8. Remove the old harness from the left rail. Be sure to save the bushing for use in installing the new harness.
5.5.5.3 Reinstalling the New Harness on the Left Rail
1. On the left rail end of the replacement harness, mark the location on the harness where the cable tie was
located.
Cable Tie
Figure 5-37
Cable Tie
2. Remove the cable tie that secures the harness to the carrier.
3. Open the first 4 cable guard latches.
4. Unsnap and remove the end link from the cable carrier.
5. Thread cable through the left rail and out hole on side
6. Pull cable through the rail until the cable mark is in the correct position just inside rail.
7. Reinstall the bushing in the hole on the side of the rail
8. Verify the mark on the cable is still in the correct position and reconnect the cable carrier to the end link
riveted to the left rail. Close the other 4 cable guard latches.
9. Reinstall new cable tie to secure the harness to the carrier
10. Reinstall the 3-pin connector.
108
6600-0343-000 104
© 2001 by Datex-Ohmeda, Inc.. All rights reserved.
Chapter 5: Repair Procedures
Содержание Giraffe OmniBed
Страница 10: ...x 6600 0343 000 104 2001 by Datex Ohmeda Inc All rights reserved Table of Contents ...
Страница 28: ...xxviii 6600 0343 000 104 2001 by Datex Ohmeda Inc All rights reserved About this Manual ...
Страница 38: ...10 6600 0343 000 104 2001 by Datex Ohmeda Inc All rights reserved Chapter 1 Functional Description ...
Страница 48: ...20 6600 0343 000 104 2001 by Datex Ohmeda Inc All rights reserved Chapter 2 Service Checkout ...
Страница 56: ...28 6600 0343 000 104 2001 by Datex Ohmeda Inc All rights reserved Chapter 3 Calibration and Maintenance ...
Страница 260: ...232 6600 0343 000 104 2001 by Datex Ohmeda Inc All rights reserved Chapter 6 Illustrated Parts ...