4.3 Alarm Messages
This section describes the alarm messages that may appear for the Giraffe OmniBed. The alarms have been
organized in this section alphabetically.
4.3.1 “Air Probe Failure” Alarm
Cause or Circumstances
Failure
Action
The software detected a
difference in the two air
thermistors of > 0.5ºC
Defective compartment air
probe or harness
Measure the resistance of the two sensors at the
harness from control board J1, Pins 1-2, and 3-4. Under
stable conditions, the resistances should be within
about 100 ohms. If defective, disassemble the air probe
assembly and disconnect the 5 pin connector inside.
Measure resistance of the two thermistors at pins
1-2 and 3-4 to determine if the sensor or harness is
defective. Refer to the Tips section (4.6.7) for resistance
values.
Defective control board.
If the thermistors and harness measure OK, replace the
control board.
4.3.2 “Air Temp >38C” or “Air Temp >40C” Alarm
Cause or Circumstances
Failure
Action
Air Temp >38C or Air Temp
>40C alarm activated even
though compartment temp is
< alarm temp.
This alarm does not self
reset after alarm condition is
resolved.
None
Push the alarm silence button to clear the alarm.
4.3.3 “Baby Hot” or “Baby Cold” Alarm
Cause or Circumstances
Failure
Action
Baby Hot or Baby Cold alarm
activated while unit is in
Manual or Air mode.
MANUAL TEMP alarm is set
as active on the nurse setup
screen or the service screen.
None
On service screen or nurse setup screen, select OFF for
the MAN TEMP ALARM setting.
34
6600-0343-000 104
© 2001 by Datex-Ohmeda, Inc.. All rights reserved.
Chapter 4: Troubleshooting
Содержание Giraffe OmniBed
Страница 10: ...x 6600 0343 000 104 2001 by Datex Ohmeda Inc All rights reserved Table of Contents ...
Страница 28: ...xxviii 6600 0343 000 104 2001 by Datex Ohmeda Inc All rights reserved About this Manual ...
Страница 38: ...10 6600 0343 000 104 2001 by Datex Ohmeda Inc All rights reserved Chapter 1 Functional Description ...
Страница 48: ...20 6600 0343 000 104 2001 by Datex Ohmeda Inc All rights reserved Chapter 2 Service Checkout ...
Страница 56: ...28 6600 0343 000 104 2001 by Datex Ohmeda Inc All rights reserved Chapter 3 Calibration and Maintenance ...
Страница 260: ...232 6600 0343 000 104 2001 by Datex Ohmeda Inc All rights reserved Chapter 6 Illustrated Parts ...