5.5 Left Rail Internal Repairs
5.5.1 Removing the Inner Rail Assembly
Refer to “Figure 5-32 Left Inner Rail”.
STEP 7
Lift rail
Cable carrier
Cable carrier channel
Left upright
Side roller bracket
Figure 5-32
Left Inner Rail
1. Using a 2.5 mm hex key, remove the 2 screws in the rear dovetail of the upright that secure the cable
carrier channel to the inside of the upright. One screw is located at the top of the upright; the other is half
way down. Using a 3mm hex key, remove the side roller bracket. Pull the entire rail assembly out of the
upright.
2. Remove the pinch guard by sliding it up then pulling it off.
102
6600-0343-000 104
© 2001 by Datex-Ohmeda, Inc.. All rights reserved.
Chapter 5: Repair Procedures