4.3.4 “Baby Mode Disabled” Alarm
Cause or Circumstances
Failure
Action
Baby mode has been disabled
in the service screen and
baby mode button is pressed.
None
Press alarm silence button to deactivate the alarm.
If you wish to use the baby mode, re-enable it in service
screen.
4.3.5 “Baby Probe 1 Failure” or “Baby Probe 2 Failure” Alarm
Cause or Circumstances
Failure
Action
Probe thermistors show
difference > 0.5ºC.
Defective patient probe
Replace patient skin probe.
Defective patient probe jack
wiring
Plug patient probe into probe jack 1. In service mode,
note the difference between readings from P11 and
P12. Plug the same probe into probe jack 2 and note the
difference between readings from P21 and P22. If they
are greater there is resistance in the probe jack harness.
Replace the harness.
4.3.6 “Bad Membrane Switch” Alarm
Cause or Circumstances
Failure
Action
All
Defective membrane switch
panel
Test switches in service mode. Replace membrane
switch panel.
© 2001 by Datex-Ohmeda, Inc.. All rights reserved.
6600-0343-000 104
35
Chapter 4: Troubleshooting
Содержание Giraffe OmniBed
Страница 10: ...x 6600 0343 000 104 2001 by Datex Ohmeda Inc All rights reserved Table of Contents ...
Страница 28: ...xxviii 6600 0343 000 104 2001 by Datex Ohmeda Inc All rights reserved About this Manual ...
Страница 38: ...10 6600 0343 000 104 2001 by Datex Ohmeda Inc All rights reserved Chapter 1 Functional Description ...
Страница 48: ...20 6600 0343 000 104 2001 by Datex Ohmeda Inc All rights reserved Chapter 2 Service Checkout ...
Страница 56: ...28 6600 0343 000 104 2001 by Datex Ohmeda Inc All rights reserved Chapter 3 Calibration and Maintenance ...
Страница 260: ...232 6600 0343 000 104 2001 by Datex Ohmeda Inc All rights reserved Chapter 6 Illustrated Parts ...