4.2 Service Screen
To access the service screen, hold in the override button (>37) during power up until the software revision
screen appears. Release the button and the first service screen will appear.
Figure 4-1
First Service Screen
Default options that may be selected from this screen appear along the left side of the screen
DAC Volt
Digital/Analog Converter voltage. This is the over temperature voltage that is used by the system to verify
the computer independent circuitry is working. To manually test this circuit enter voltages from 0 to .5 V.
The DV value at the bottom right corner of the screen should match this value within 10mV.
Language
English is the default language that appears on the EL display, but you can select French, Spanish, etc.,
depending on what software is installed.
Temp U
Changes temperature displays to show “Fahrenheit”, “Celsius” or “Celsius Only” so the Fahrenheit option
is not present on the user Set-up screen (Celsius is factory set default).
Volume
Select one of four volume settings. 1 is minimum, 4 is maximum.
Pat Alarm
Set the default Hot Baby/Cold Baby alarm to activate when either 0.5ºC or 1.0ºC difference is read
between a set temperature and the baby probe temperature.
Elevate
Disable or enable the bed up/down pedal switches. If the pedals are disabled on the service screen, they
cannot be enabled on the user setup screen (wrench icon).
Pat Ctrl
Allows you to disable patient control.
Pat Algo
To be used for future software options.
Preheat
Select from 10 to 50% radiant heater power to preheat without alarms; 25% is the default.
NOTE:
Resetting maximum preheat level to above 25% will result in noncompliance to device standard
IEC 60601-2-21
Canopy
Disable or enable the canopy pedal switches; hand switches will remain active. If the switches are
disabled on the service screen, they cannot be enabled on the user setup screen (wrench icon).
Scale U
Select from “Grams”, “Pounds” or “Grams Only” so that the pounds option is not present on the user Scale
screen (grams is the factory default).
30
6600-0343-000 104
© 2001 by Datex-Ohmeda, Inc.. All rights reserved.
Chapter 4: Troubleshooting
Содержание Giraffe OmniBed
Страница 10: ...x 6600 0343 000 104 2001 by Datex Ohmeda Inc All rights reserved Table of Contents ...
Страница 28: ...xxviii 6600 0343 000 104 2001 by Datex Ohmeda Inc All rights reserved About this Manual ...
Страница 38: ...10 6600 0343 000 104 2001 by Datex Ohmeda Inc All rights reserved Chapter 1 Functional Description ...
Страница 48: ...20 6600 0343 000 104 2001 by Datex Ohmeda Inc All rights reserved Chapter 2 Service Checkout ...
Страница 56: ...28 6600 0343 000 104 2001 by Datex Ohmeda Inc All rights reserved Chapter 3 Calibration and Maintenance ...
Страница 260: ...232 6600 0343 000 104 2001 by Datex Ohmeda Inc All rights reserved Chapter 6 Illustrated Parts ...