4.3.10 “Canopy Pedal Disabled” Alarm
Cause or Circumstances
Failure
Action
Canopy pedal switches have
been disabled in nurse setup
screen or service screen and
pedal is pressed.
None
When pedal is released, the alarm deactivates.
NOTE:
If the canopy pedal switches are disabled in
the service screen and you wish to use them, you can
reactivate them only in the service screen.
4.3.11 “Canopy Pedal Failure” Alarm
Cause or Circumstances
Failure
Action
One of the canopy pedal
switches is shorted. (The
alarm appears about 100
seconds after the short.)
Defective canopy pedal
switch or harness.
Check the switches on the service screen to determine
which is shorted. Refer to the Tips section (4.6.9). Raise
the lower shroud and check continuity at the switch
connectors to determine if a switch is defective or the
wiring has a short.
One of the canopy hand
switches is shorted.
Defective canopy hand switch
or harness.
Replace one switch. If the error persists, replace the
other switch.
4.3.12 “Canopy Pedal Pressed” Alarm
Cause or Circumstances
Failure
Action
The canopy pedal switch was
pressed and held on during
power up.
None
When the canopy pedal switch is released, the alarm
deactivates. If alarm silence button is pressed or the
pedal remains on for 100 seconds the Canopy Pedal
Failure alarm activates and the canopy pedal switches
are disabled. You can still raise and lower the canopy
using the hand switches. Shut down the unit and power
up to clear failure.
Other
Canopy pedal switch has
failed
Refer to section 4.3.11 for troubleshooting information.
4.3.13 “Down Pedal Pressed” Alarm
Cause or Circumstances
Failure
Action
The bed down pedal switch
was pressed and held on
during power up.
None
When the bed down pedal switch is released, the
alarm deactivates. If alarm silence button is pressed or
the pedal remains on for 100 seconds the Bed Down
Pedal Failure alarm activates and the bed down pedal
switches are disabled. Shut down the unit and power up
to clear failure.
Other
Bed down pedal switch has
failed
Refer to section 4.3.8 for troubleshooting information.
© 2001 by Datex-Ohmeda, Inc.. All rights reserved.
6600-0343-000 104
37
Chapter 4: Troubleshooting
Содержание Giraffe OmniBed
Страница 10: ...x 6600 0343 000 104 2001 by Datex Ohmeda Inc All rights reserved Table of Contents ...
Страница 28: ...xxviii 6600 0343 000 104 2001 by Datex Ohmeda Inc All rights reserved About this Manual ...
Страница 38: ...10 6600 0343 000 104 2001 by Datex Ohmeda Inc All rights reserved Chapter 1 Functional Description ...
Страница 48: ...20 6600 0343 000 104 2001 by Datex Ohmeda Inc All rights reserved Chapter 2 Service Checkout ...
Страница 56: ...28 6600 0343 000 104 2001 by Datex Ohmeda Inc All rights reserved Chapter 3 Calibration and Maintenance ...
Страница 260: ...232 6600 0343 000 104 2001 by Datex Ohmeda Inc All rights reserved Chapter 6 Illustrated Parts ...