Action
Step
Edit the file name by using the Right arrow key to move into the file name,
and Up and Down arrows to adjust the characters as desired. To skip editing
the name use the Left arrow to proceed to the next line.
3
To change any letter, move the cursor to that location using the Right and
Left arrow keys and then scroll through the 37 available options using the
Up and Down arrow keys. Once the desired letter appears, move the cursor
to the next character to be set.
The original label “File #” occupies 6 of the 16 available characters. To change
the label to a more descriptive name, such as “11 CM 3 to 10L,” requires
scrolling through the alphanumeric options for each of the 11 digits in the
name.
P r o t # 1 F i l e 1
R e h y d r a t e 1 0 : 0 0 H r @ 2 0
˚
C
I E F @ 2 0
˚
C 5 0
µ
A / S t r i p
1 S t r i p s 0 S t e p s
Edit time range by using the Left and Right arrow keys to position the cursor,
and use the Up and Down arrow keys to set the time range.
4
The available time range is 0–99:59 h, with 10 min resolution. (Hence, only
the first three digits are editable). A minimum of 10 hrs is required for the
rehydration of IPG strip gels, but overnight is recommended. This protocol
step can also be used as a simple timer to start the protocol at a convenient
time. If the IPG strips are already rehydrated and ready for immediate focus-
ing, set the rehydration duration to 00:00 Hrs.
P r o t # 1 F i l e 1
R e h y d r a t e 1 0 : 0 0 H r @ 2 0
˚
C
I E F @ 2 0
˚
C 5 0
µ
A / S t r i p
1 S t r i p s 0 S t e p s
76
Ettan IPGphor 3 Operating Instructions 28-9635-51 AC
5 Operation
5.3 Perform a run from the control panel
5.3.3 Set protocol parameters