WARNING
Do not block the rear and side panel of Ettan IPGphor 3. The power
switch must always be easy to access. The power cord must always
be easy to disconnect.
Two power cords are supplied: one cord has a “North American” style (UL817) 3-pronged
plug for 115 V AC power outlets, and the other has a “Central European” style (CEE7/VII)
plug for 230 V AC power outlets.
Action
Step
Select the cord with the appropriate plug for the power outlet in your labo-
ratory. If neither of these cords are suitable for your power outlets, obtain
a detachable cord with an IEC/320/C13 (CEE22/V) outlet.
1
Plug the connectors into the power cord receptacle on the rear panel and
into a properly grounded power outlet.
2
Turn on power and run
diagnostic program
Turn on the instrument and run a diagnostic program according to the description in
Section 5.2 Start up the instrument, on page 66.
Serial port connector for a
computer connection
NOTICE
Any computer used with the equipment shall comply with IEC 60950
and be installed and used according to the manufacturer's instruc-
tions.
The RS232 serial port for a computer connection is placed at the rear of Ettan IPGphor
3. The type of cable required (shielded straight serial or shielded null modem) depends
on the type of device (DTE or DCE) that it is connected to:
•
If the computer is configured to receive data on pin 2 and transmit on pin 3, a regular
straight serial cable is required.
•
For any other configuration, a null modem cable is required.
42
Ettan IPGphor 3 Operating Instructions 28-9635-51 AC
4 Installation
4.3 Setup