Position the electrodes
Follow the instructions below to position the electrodes.
Action
Step
With the electrode cams in the open position, position the electrode assembly
on top of all the wicks. The electrode must be in contact with the wick.
1
Swivel the cams into the closed position under the external lip of the tray.
The electrodes should not be moved while the cams are in the closed position.
2
64
Ettan IPGphor 3 Operating Instructions 28-9635-51 AC
5 Operation
5.1 Preparations before start
5.1.2 Use the Ettan IPGphor Cup Loading Manifold